Total Chlorine
50081_Chlorine_Total_Forward.fm
Page 105
Total Chlorine
• To avoid loss of chlorine, be careful not to agitate the sample when measuring or pouring.
• Sample glassware must be cleaned and pretreated to satisfy any chlorine demand before sample is introduced. See
• Always use organic-free reagent water for sample dilution (Cat. No. 26415-49).
• When sampling tap water, let the faucet run for at least 4–5 minutes prior to collecting the sample.
• Prepare a test sample diluting Chlorine Standard Solution (Cat. No. 14268-10) with deionized (DI) water.
• Rinse the electrode, temperature probe, and anti-diffusion tip with deionized water before every titration.
• Download the “Certificate of Analysis” (COA) to obtain the exact concentration of any unopened bottle of Hach titrant or standard
solution. See
Obtaining the Certificate of Analysis for Hach Reagents
.
• Hach buffer reagents for chlorine titrations are highly recommended for this analysis.
• Never substitute buffers designed for calibrating pH meters. They may contain dyes that interfere with amperometric titration.
• Never use buffers contaminated with mold or bacteria.
• Adjust the contrast of the display by accessing the Methods menu, then pressing
7
to lighten,
0
to darken.
• A distinction is drawn between running a new test and a new sample. Each test is a replicate run of the current, or previous,
sample analysis. The AutoCAT automatically tracks the results of a series of tests, and automatically calculates the mean and
standard deviation for all the results. Each test is given a different number, all of which are listed under the same Sample ID.
When starting a new sample, a new ID is assigned.
• The AutoCAT can accommodate any sample size. The default sample volume of 200 mL is recommended in
Standard Methods
,
and can be adjusted to expand the test range.
• The AutoCAT calculates the chlorine concentration based on the sample volume. Make sure that the sample volume is correct.
• The analysis data may be analyzed as many times as needed from the Zoom Range Determination window, in manual (MEPD)
and/or automatic (AEPD) modes. The results are not stored until accepted by the user.
• The
STOP
key can be pressed at any time to interrupt instrument operation. The user will be given the option to resume, restart,
or quit completely.
• Press the
CHECK MARK
key at any time during data acquisition to halt data collection.
• Press the
DEL
button during any timer period except the Start Timer to immediately exit the timer and proceed to the next step in
the procedure.
• Flush the burette each day before the first sample test or calibration is performed. See Daily Start-up on page 26.
• Flush the burette when changing titrants. See Changing Titrants on page 25.
Amperometric Forward Titration
Titration Workstation
0.0012–5.000 mg/L as Cl
2
*
Scope and Application:
For water and wastewater; USEPA accepted for reporting**
* Higher ranges can be determined through sample dilution. See
.
**Procedure equivalent to USEPA methods 330.1 and 330.3, and
Standard Methods
4500-Cl D. for wastewater.
Sample Tips and Techniques
Reagent Tips and Techniques
Instrument Tips and Techniques
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......