Sulfite
Sulfite
50081_Sulfite_Back.fm
Page 115
17.
The Zoom Range
Determination window will
display the titration curve.
Use the navigation keys to
select and move the cursors
(vertical lines) to define the
range of data points for
analysis.
Proceed to
.
Note:
Defining the range of
data points allows the user to
eliminate undesirable portions
of the titration curve from the
analysis.
18.
Select the type of end
point determination.
Press the
CHECK MARK
for
“Automatic” and proceed to
or press
DEL
for
“Manual”.
Note:
Automatic determination
attempts to identify linear
segments without user input.
Note:
This screen is not
displayed if the Auto-detection
software is enabled.
19.
Select two line
segments, one on each side
of the titration end point, by
positioning the cursors
(vertical lines) at the
beginning and end of each
segment.
Press the
CHECK MARK
to
confirm changes or press
DEL
to return to the previous
window.
Note:
For more information on
manual EP determination, see
page 135.
20.
Press the
CHECK MARK
to accept the analysis.
Proceed to
or press
DEL
to return to the previous
window (Zoom Range
Determination). Adjust the
linear segments and
reanalyze the data.
21.
The AutoCAT 9000 will
run the analysis based on
the automatic selection of
linear segments.
Note:
The automated analysis
requires approximately
1 minute to complete.
22.
Press the
CHECK MARK
to accept the analysis. Press
DEL
to return to the previous
window (Zoom Range
Determination) to adjust the
linear segments and
reanalyze the data.
23.
The analysis for this test
is complete.
Press
1
to select Continue
and perform additional tests
or finish analyses. Press
2
to
select More Details and view
analysis statistics for the
current Sample ID.
Note:
The results show the
concentration of one of these
analytes, depending on what
form of the test is being
performed. The results given
do not indicate the presence of
all compounds listed.
24.
Press
1
to select
New Test and run a replicate
sample.
Press
2
to select New Sample
and initiate analysis of a new
sample with a new Sample
ID.
Press
3
to select
End Analysis and return to
the Methods menu.
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......