![Hach AutoCAT 9000 Instruction Manual Download Page 45](http://html1.mh-extra.com/html/hach/autocat-9000/autocat-9000_instruction-manual_3154724045.webp)
Collecting and Accessing Data
50081_AutomatedTitration.fm
Page 43
Section 5
Collecting and Accessing Data
5.1 Using the AutoCAT 9000 for Amperometric Methods
The following analysis procedures, based on Standard Methods, are embedded directly
in the instrument software:
•
Forward titration for free chlorine
•
Forward titration for total chlorine
•
Back titration for total chlorine
•
Forward titration for chlorine dioxide
Hach sells pre-standardized titrants, prepared according to Standard Methods, for both
forward and back titrations. The 0.00564 N phenylarsine oxide titrant (Cat. No. 1999-53)
has a shelf life of >1 year in the unopened bottle. The unopened 0.0282 N Iodine
Standard Solution titrant (Cat. No. 23333-53) is stable for 6 to 12 months.
Periodically recalibrate the titrants after opening to ensure accuracy. Dispense reagents
directly from the product bottle to eliminate contamination from transferring them to
separate containers.
5.2 AutoCAT Titration
AutoCAT 9000 functions can be separated into three categories: (1) data collection,
(2) end point determination, and (3) calculation. Each category is described in the
sections below.
5.2.1 Data Collection
The AutoCAT displays the titration curve (current (µA) vs. titrant volume (mL))
graphically in real time, as it dispenses titrant and measures the current. After the
titration is complete, the user can choose between a fully automated or manual routine to
determine the end point. Following end point determination, the AutoCAT 9000
automatically calculates the analyte concentration.
The titration curve is constructed by plotting the measured current as a function of the
volume of titrant added. Several parameters, which must be set prior to data collection,
are available to control the data collection process. These parameters include the
(1) current range, (2) volume increment, (3) predose volume, and (4) Auto-scaling.
5.2.1.1 Current Range
Current Range allows the user to set the maximum current (in µA) expected for a
particular titration. The AutoCAT 9000 allows four possible settings (2 µA, 20 µA,
200 µA, and 1 mA) for the current range. The default setting (20 µA) setting is adequate
for most purposes. The 2 µA setting is recommended when analyzing samples with
concentration less than 0.600 mg/L Total Chlorine.
Note:
The 200-µA and 1-mA settings are available for experimental use, and are not
recommended for most titrations.
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......