AutoCAT 9000 Validation Study
AutoCAT 9000 Validation Study
Page 162
50081_AutoCAT 9000 Validation Study.fm
A comparison graph at a chlorine concentration of 16 µL Cl
2
is shown in
.
Figure 2
Titration Curve—16 µg/L Chlorine
Spreadsheet analysis of the data pairs, utilizing linear regressions of the line segments
prior to and after the endpoint, and results calculation, resulted in equivalent results to
those displayed on the instrument, again with excellent correlation and signal to noise.
MDL, Accuracy, and Precision Determination
After these trials were concluded, demonstrating the instrument capability to measure
low µg/L chlorine concentrations, per Standard Method 4500 Cl D, experiments were
conducted to determine the Method Detection Limit (MDL), according to 40 CFR Part
136, Appendix B.
This MDL calculation involves replicate analyses, where the sample concentration must
be within the range 1 to 5 times the calculated MDL. The MDL is calculated as the
Student's t value (at the 99% confidence interval), times the standard deviation. For
seven replicates, the Student's t value (at the 99% confidence interval) is 3.143.
The experimental conditions, and apparatus for this series of experiments was the same
as for the screening studies described above. Great care was observed during the required
sample manipulations to minimize experimental error.
Seven replicate measurements were initially run, at a sample chlorine concentration of
6.5 µg/L. An analysis of the standard deviation resulting from these trials determined that
the sample concentration was too high, outside the range 1 to 5 times the calculated
MDL. A second set of seven replicate measurements was then conducted on a chlorine
sample concentration of 4.1 µg/L. An analysis of the calculated MDL from this series of
trials showed that the sample concentration was within the stipulated range.
The resulting MDL, in accordance with 40 CFR, Part 136, Appendix B, was thus
determined to be 1.2 µg/L using this instrument. A summary of these trials can be seen in
.
Table 1 Method Detection Limit
Method
Standard
Standard Deviation
Student t
Calculated MDL
4500 Cl D
6.5 µg/L
±0.40 µg/L
3.143
1.3 µg/L
4500 Cl D
4.1 µg/L
±0.37 µg/L
3.143
1.2 µg/L
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......