Basic Operation
Basic Operation
50081_Operation.fm
Page 25
•
Method library—
(available in Supervisor Mode) contains all of the adjustable
method parameters. Those with supervisor access can edit, reset, and delete method
parameters to suit specific needs.
•
GLP-Archives
—provides access to the Good Laboratory Practice (GLP) tables, the
stored method analysis results, and global variables.
•
Method Tab
—is an animated icon that becomes active when a titration/calibration is
being performed.
Note:
From the Methods menu, adjust the contrast of the display by pressing
7
to lighten,
and
0
to darken.
3.2.3 Titrants Menu
The Titrants menu provides access to the embedded calibration procedures, adjustable
parameters, stored calibration results, and burette functions.
•
Title bar
—displays Titrants
•
Calibrate/Enter titrant conc.
—determines the concentration of the titrant by
running a calibration or calibration sequence. Alternatively, the concentration can be
input manually on the alphanumeric keypad.
•
Install titrants
—allows the user to install or replace the titrants in the selected
method.
•
Check titrants
—allows the user to verify the parameters of the titrant used in the
selected method.
•
Titrant library
—contains all of the adjustable calibration parameters. Supervisors
can choose to have the titrant concentration input manually or determined through
embedded calibration.
•
GLP-Archives
—provides access to the Good Laboratory Practice (GLP) tables and
display the stored titrant calibration results.
•
Burette functions
—allows the user to fill, empty, flush, rinse, and replace the
burette.
•
Titrants tab
—displays one of four titrant status icons, see
.
3.2.3.1 Changing Titrants
The auto-flush routine takes 60 seconds and uses 10–12 mL of titrant. This function
provides a thorough flush and should be used when changing to a different titrant.
1.
From the Titrant menu, press
6
to access Burette Functions.
2.
Press
3
to begin the flush cycle. Allow the cycle to run to completion.
Table 1 Titrant Status Icons
Sunny Icon—the titrant calibration or manual entry of the titrant concentration has
been performed satisfactorily for the current titrant.
Cloudy Icon—either the titrant calibration should be performed within 24 hours, or
the titrant will expire in less than one week.
Stormy Icon—the titrant calibration or the titrant in the system has expired, the
calibration has failed to satisfy the user-programmed acceptance limits, or the
installed titrant and selected method are incompatible.
Question Mark—a problem exists within the method parameters for the titrant.
Supervisory level access is required.
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......