SILKOSTAT 9210 - MANUEL D'UTILISATION
Typical measurement values during calibration
These values are given as an indication and are dependant on the ambient temperature. The real values
may deviate from more than one volt without having an effect on the measurements.
Version 0…1000 ppb
Sample concentration
Reference (mV)
Measure (mV)
107 ppb
3500
2515
214,9 ppb
3500
1754
267,4 ppb
3500
1472
534,8 ppb
3500
603
Version 0…5000 ppb
Sample concentration
Reference (mV)
Measure (mV)
1000 ppb
3500
1400
3000
ppb 3500 300