Page 71
SILKOSTAT 9210
Prolonged stopping
In case of an extended stop longer than 1 week, it is recommended
to clean the tubes and the measuring cell.
Provide 250 ml of demineralised water.
Submerge the 3 tubes in the beaker and start the rinsing operation:
•
Stop the instrument by pressing "STOP".
•
Press the "MENU" key.
EXTENDED STOP
Put tubes in
demineralised water
Accept
with
OK
OK
Esc
•
Choose the "MAINTENANCE" menu.
•
Choose the sub menu "EXTENDED STOP".
The measuring cell is empty.
Disconnect the instrument.
Replace the reagents tubes in their canisters.
In case of an extended stop over 6 weeks, you must throw the
reagents and prepare another set for the restart.
Note : Don’t reverse the tubes.