Page 38
SILKOSTAT 9210
Execution primary sub menu
CALIBRATION
PROGRAMMING
EXECUTION MANUAL CAL
PARAMETERS
HISTORIC
Select
Esc
A primary calibration is carried out when operating the instrument
for the first time and is used as a reference for the following
calibrations. Generally, it should be carried out only after a
consequent modification.
It allows to fix the zero and slope reference.
All the results of the further manual or automatic calibrations will
be compared to the zero and slope references set by this
calibration.
A more important shift will allow to detect an error calibration
which will be signified by an error message.
This calibration can be realized only in the manual way.
•
Access to the main "MENU".
•
Choose "CALIBRATION" menu with the "select" function key
and validate by pressing "Enter".
•
Choose "EXECUTION PRIMARY" sub menu and validate by
pressing "Enter".
The slope calibration is going on, followed by the zero.
PRIMARY CALIBRATION
Slope
calibration
Cal sol sampling
CALIBRATION
Esc
The upper bar indicates the type of calibration : primary, manual
or automatic.
Press "Esc" if you want to quit the calibration menu : and go back
to the "CALIBRATION" menu.
The calibration stages are displayed as follow :
•
Sample
preparation.
•
Conditioning
•
Measure.
•
Rinse.
A bargraph indicates the calibration state.
At the end of the calibration, results are displayed on screen.
EXECUTION PRIMARY CAL
Summary of Contents for POLYMETRON 9210
Page 1: ...Operator Manual POLYMETRON 9210 221 192 010 Revision J 20 11 2007...
Page 2: ......
Page 10: ......
Page 11: ...Page 5 SILKOSTAT 9210 1 Presentation of the instrument Figure 1...
Page 18: ......
Page 28: ......
Page 74: ......
Page 80: ......
Page 114: ......
Page 115: ......
Page 116: ......