SILKOSTAT 9210
Calibration menu
CALIBRATION
PROGRAMMING
EXECUTION PRIMARY CAL
EXECUTION MANUAL CAL
PARAMETERS
HISTORIC
Select
Esc
To get a good precision from the instrument, you should calibrate
it on a regular basis.
Calibrations allow to adjust :
•
ZERO of the system.
•
SLOPE of the system.
•
Both ZERO and SLOPE of the system.
The ZERO calibration is performed by the
SILKOSTAT/chemically.
In order to avoid the use of water free of SiO2, the SILKOSTAT
carries out a measurement without a colorimetric reaction.
The slope of the system is calibrated with a known concentration
of SiO
2
standard solution.
Three types of calibration are available :
•
Primary
calibration.
•
Manual
execution.
•
Automatic
execution.
MENU
CALIBRATION
MAINTENANCE
MEASURE
ALARMS
mA
OUTPUTS
COMMUNICATION
SEQUENCE
SERVICE
Select
Esc
"MENU" allows to access to the programming of the instrument.
This screen allows to choose between different programming
sub-menus.
Press the "Select" function key to select the programming menu
you wish to have access to and press "Enter" to validate the
choice.