Page 54
SILKOSTAT 9210
Alarms Menu
Alarms parameters 1 to 6 sub menu
Value
Description
MODE
−
Limit
−
Active
channel
−
Level
sample.
Allows to choose the type of relay : threshold (limit) , active channel
or level sample.
AFFEC
Channel 1 to 6 Allows to assign the alarm (1 to 6) to the measure channel of our
choice.
LIMIT
XXXX
Allows to adjust the limit value (ppb)
DIRECTION
−
Up
−
Down
Allows to adjust the relay direction interlocking
HYSTERESIS
XX %
Allows to adjust the hysteresis in %
RELAYS N.O
N.C
Relay type
N.O : normally open
N.C : normally closed
ALARMS
ALARM 1
ALARM 2
ALARM 3
ALARM 4
ALARM 5
ALARM 6
WARNING ALARM
SYSTEM ALARM
Select
Esc
This screen allows to assign alarms and parameters to each relay.