Page 46
SILKOSTAT 9210
Reagents refill sub menu
Adjust volumes
START UP
Pump priming
OK
Esc
Pump priming :
Press OK when the pump tubing are full of reagent. The display
goes back to the menu “Maintenance”.
REAGENTS REFILL
ADJUST VOLUMES
TUBING DRAINING
CANISTERS FULL yes/no
Select
Esc
Every reagent and calibration solution renewal, access this menu
for.
This menu allows to choose between :
ADJUST VOLUMES :
the 4 canisters volumes are displayed and
can be modified.
TUBES PRIMING :
start the priming / renewal procedure of the
reagents in the tubes. "Esc" allows to return to the previous
screen or to stop the procedure.
CANISTERS FULL :
choose this option each time than canisters
are full. The volumes of the 4 canisters are automatically
programmed with 2 litres.
ADJUST VOLUMES
REAGENT VOL 1 : 2.00
λ
REAGENT VOL 2 : 2.00
λ
REAGENT VOL 3 : 2.00
λ
REAGENT VOL: 2.00
λ
Select
This menu allows to visualize the volumes of the reagent canisters
and the calibration solution. You eventually can adjust these
volumes. The «FULL CANISTERS » menu put these volumes to
2 litres.
The analyser controls the calibration solution and reagent
consumption and emits alarm system according to the level of the
canisters.
Warning
Alarm system
Reagents
< 250ml
< 150ml
Esc
Cal solution
< 300ml
Summary of Contents for POLYMETRON 9210
Page 1: ...Operator Manual POLYMETRON 9210 221 192 010 Revision J 20 11 2007...
Page 2: ......
Page 10: ......
Page 11: ...Page 5 SILKOSTAT 9210 1 Presentation of the instrument Figure 1...
Page 18: ......
Page 28: ......
Page 74: ......
Page 80: ......
Page 114: ......
Page 115: ......
Page 116: ......