Page 43
SILKOSTAT 9210
Manual – automatic calibration results sub menu
At the end of a manual or automatic calibration, the analyser compares
the results with the primary calibration master.
According to the calibration results, you can validate, restart or quit the
calibration.
There are 3 zones :
S : (%)
20 %
50 %
150 %
180 %
Rejection zone
Confirmation
zone
Zone ok
Confirmation
zone
Rejection zone
O : (mAbs)
- 50
- 30
30
50
XXX
Primary calibration :
S = XXX XX:XX:XX
O = XXX XX:XX:XX
Last
calibration
S = XXX XX:XX:XX
O = XXX XX:XX:XX
C = XXX XX:XX:XX
Actual calibration
S = XXX
O = XXX
Calibration OK
OK
Esc
Zone ok
Normal variations around the reference values. The new
parameters are considered.
In automatic mode, the analyser returns to its measurement cycle.
In manual mode, validate with OK and restart the instrument
(START).
P:XXXXX DD:MM:YY slope
Z:XXXXX DD:MM:YY offset
C:XXXXX DD:MM: concentration of the sample solution just
measured, calculated using the calibration values from the "Last
calibration".