SILKOSTAT 9210
Definition of the display screen 3 (Disp3) : alarms
ALARM
S1 : OK
.
.
S6 : OK
W : disable
:
OK
DISP4
Esc
This screen displays the alarms status as well as the following
indications :
•
S1…S6
:
Threshold function with the following
parameters,
* OK :
Alarm condition is not met.
* Sample X > XXX : Alarm is actived and the alarm condition is
clearly mentioned. For example, the
sample X measurement exceeds xxx ppb.
* Disable:
Alarm function is disactivated.
•
W :
Warning
alarm.
•
:
System
alarm.
•
DISP4 :
Go to the display 4.
The following table indicates the different messages which may appear in
case of default.
Displayed message
(english)
Description Category
Reset mod. Measure
Reset of the measure module
System
Reset module LC
Reset of the local controller module.
System
Error lc
Local controller module error
System
Err. measure mod.
Measure module error
System
Com error
CAN communication error
System
Table error
Error during table transmission
System
Task error
Task execution error
System
Measure error
Measurement error
Warning
Reference error
LED light ajustment problem
Warning
Extinction error
Black current problem
Warning
Meas.time over
Maximum cycle time for the measure
has been exceeded
Warning
Cal.time over
Maximum cycle time for calibration has
been exceeded
Warning
No sample
No sample available
System
Sample x missing
Sample x missing
Warning
Reagent < 250 ml
Low level of the reagent left (2.5 days
before no reagent alarm)
Warning
Calib < 300 ml
Low level of the calibration solution (1
calibration left before no calibration
solution alarm)
Warning
No reagent
Empty reagent canisters <150 ml
System
Slope drift
Slight drift of slope calibration
parameter (Confirmation zone)
Warning
Zero drift
Slight drift of zero offset calibration
parameter (Confirmation zone)
Warning
Zero error
Zero calibration major error
System
Slope error
Slope calibration major error
System
Sample x < C
Sample concentration x is lower than
C
Threshold
Sample x < 0 ppb
Sample concentration x is negative
Warning
Sample x > 1000 ppb ou >
5000 ppb
Sample concentration over range
Warning
Reference error
The reference measurement lies
outside its limiting values
Warning
.
.
.
.
Summary of Contents for POLYMETRON 9210
Page 1: ...Operator Manual POLYMETRON 9210 221 192 010 Revision J 20 11 2007...
Page 2: ......
Page 10: ......
Page 11: ...Page 5 SILKOSTAT 9210 1 Presentation of the instrument Figure 1...
Page 18: ......
Page 28: ......
Page 74: ......
Page 80: ......
Page 114: ......
Page 115: ......
Page 116: ......