Page 61
SILKOSTAT 9210
Communication menu
Choice of the measure sequence menu
Channel activation sub menu
COMMUNICATION
N : 00
BAUD : 9600
PARITY : No
STOP BIT : 1
SWAP WORD : No
Select
Esc
This screen allows to adjust the communication parameters like
RS 485 of the SILKOSTAT. The communication protocol is
MODBUS-JBUS.
Utilize the
scroll up key to change the selected digit and
press the « Enter » key to validate and go to the next digit.
SEQUENCE
CHANNEL ACTIVATION
SEQUENCE
Select
Esc
The "SEQUENCE" menu allows to run or stop the different
measurement channels and to set up the analysis samples order.
CHANNEL ACTIVATION
CHANNEL 1 : On
CHANNEL 2 : Off
CHANNEL 3 : On
CHANNEL 4 : On
CHANNEL 5 : Off
CHANNEL 6 : On
Select
Esc
The « CHANNEL ACTIVATION » allows to run or to stop the
measurement channel.
Select : allows to choose the channel.
: allows to run / stop a channel
Enter : allows to validate a choice.
N° :
Monec number (0 … 33).
Baud :
Transmission speed in bauds (300, 600 1
200, 2 400, 4 800, 9 600).
Parity :
odd, even, no.
Stop bit:
1 or 2
Swap word
Allows to reverse the send order of the
data.