21
Installation
3.4.2 Physical installation
An installation wizard will start automatically once power is connected to the instrument for the
first time.
1.
Connect the instrument to a power source - see
2.
Turn the instrument
ON
. An instrument auto check routine will start automatically. As this is
the first time the instrument will have been powered up, the
O2 sensor residual
and the
Antifoam level
LEDs will always be red. This is normal as the system has not yet been fully
commissioned and these can be ignored at this stage.
3.
The wizard will start automatically. Press the
Installation
button to proceed.
4.
At the end of the process select the auto-check option. See
for a more detailed explanation of the auto-check process and results.
3.4.3 Post-installation procedures
1.
The instrument has been delivered with English as the default language. However, a
USB
memory stick
is supplied with other available languages (German, Spanish, Chinese, and
Japanese). If you wish to keep English as the preferred language please proceed to the
next step (
Step 2
below), otherwise follow these instructions for installing a new language:
a.
Make sure that ActiveSync® is installed on your PC. If you do not have it installed on
your PC, please refer to
Microsoft ActiveSync® configuration on page 22
for installation
details.
b.
Connect your PC to the instrument via the USB client connector (
No. 5
in
). ActiveSync® will start automatically. By default ActiveSync® proposes to set
up a partnership with the instrument. This is not required, so select
NO
before
continuing.
c.
Insert the supplied
USB memory stick
into your PC.
d.
Inside the “
Setup
” folder, double click on the “
setup.exe
” executable file. Follow the
on-screen instructions on your PC and then on the instrument to install the preferred
language software.
2.
Set the security levels, user ID’s and passwords as described in
.
3.
Configure the instrument parameters as described in
4.
Perform a barometric sensor calibration as described in
5.
Perform an O
2
sensor calibration as described in
. Wait until the
residual value of the sensor is low enough (controlled automatically by the instrument). This
operation may take around 1-2 hours depending on storage conditions.
Note:
The pressure, flow, temperature and CO
2
sensors are more stable and will not require calibration
prior to use.
The instrument should now be ready for use.
If a problem should arise with the instrument, please refer initially to
. If the problem cannot be overcome, please contact your Hach Lange representative
who will be happy to assist you.
Summary of Contents for ORBISPHERE 6110
Page 1: ...DOC024 52 93009 ORBISPHERE Model 6110 Package Analyzer User Manual 08 2013 Edition 5 ...
Page 2: ......
Page 6: ...4 Table of Contents ...
Page 14: ...12 General Information ...
Page 16: ...14 Specifications 2 2 Instrument dimensions Figure 1 Instrument dimensions ...
Page 28: ...26 Installation ...
Page 34: ...32 User Interface and Startup 4 3 Menu structure overview ...
Page 35: ...33 Section 5 Configuration 5 1 Configuration menu overview ...
Page 46: ...44 Configuration ...
Page 56: ...54 Measurement ...
Page 58: ...56 Analysis ...
Page 70: ...68 Spare Parts and Accessories ...
Page 74: ...72 Material Safety Data Sheets MSDS ...
Page 75: ...73 ...