20
Installation
Locally supplied
The following must be provided locally by the user:
3.3 Positioning
Hach Lange recommends installing the instrument on a sturdy laboratory table that can more
than support the 55kg weight. Place it on a clean flat surface, and convenient to the power
source. The display screen should be at head-height for easy viewing and operation.
3.4 Installation
3.4.1 Pre-installation
You will need the antifoam cartridge filled with the antifoam liquid as part of the installation
process. Prepare this before starting the installation using the instructions in
In addition make sure you have suitable standards available before performing any calibration,
see
Standards requirements on page 45
for calibrating the instrument sensors.
Your A1100 electrochemical sensor has been thoroughly cleaned and tested at the factory
before shipment and installed in the instrument. It has been shipped with a cartridge containing
a membrane and electrolyte pre-installed to protect the sensor head. The sensor must be
removed from the instrument and the cartridge removed and replaced with a new one prior to
first use to make it fully operational. These steps are described in the
A1100 cleaning
procedure for 6110 TPA
document supplied with the instrument. The new cartridge is included
in the sensor recharge kit, also supplied with the instrument.
Solenoid valve 2/2 NC with O-ring
33122
Filter with Goretex membrane (52 items)
33124
Piercing tip x1 (for PET)
DG33308
Piercing tip x2 (for metal)
33127
Piercing tip mounting tool
33159
Pneumatic tubing kit
33154
Antifoam recharge bottle 1L
33156
Antifoam cartridge (empty)
33157
Antifoam priming beaker
33161
Syringe for antifoam cartridge
33162
Sensor recharge kit
2956A-A
Outlet sample tube assembly
DG33267
USB key
60930
USB client cable including connectors, length = 3m
32533.03
Sensor support (part of A1100-S00T)
32205
PPS sensor storage and calibration cap (part of A1100-S00T)
28129
Purge gas for O
2
and CO
2
sensor (CO
2
)
6 to 7 bar (87 to 102 psia), purity > 99.9%
Forcing gas for piercing device (Air or N
2
)
5.5 to 6 bar (80 to 87 psia)
Reference gas for CO
2
sensor (optional)
5.5 to 6 bar (80 to 87 psia)
Loctite
®
222 sealant (or similar)
For locking and sealing piercing tips
Precision certified barometer
For calibrating the barometric pressure sensor
Precision certified pressure gauge (0-7 bar)
For calibrating the pressure sensor
CAUTION
The instrument is heavy (55 kg) so extreme care must be taken when handling, to avoid
damaging the instrument or inflicting personal injury. It is
highly recommended
to use a
pulley system (or similar) attached to the lifting ring located on the top of the instrument when
moving it.
Summary of Contents for ORBISPHERE 6110
Page 1: ...DOC024 52 93009 ORBISPHERE Model 6110 Package Analyzer User Manual 08 2013 Edition 5 ...
Page 2: ......
Page 6: ...4 Table of Contents ...
Page 14: ...12 General Information ...
Page 16: ...14 Specifications 2 2 Instrument dimensions Figure 1 Instrument dimensions ...
Page 28: ...26 Installation ...
Page 34: ...32 User Interface and Startup 4 3 Menu structure overview ...
Page 35: ...33 Section 5 Configuration 5 1 Configuration menu overview ...
Page 46: ...44 Configuration ...
Page 56: ...54 Measurement ...
Page 58: ...56 Analysis ...
Page 70: ...68 Spare Parts and Accessories ...
Page 74: ...72 Material Safety Data Sheets MSDS ...
Page 75: ...73 ...