25
Installation
7.
Attach the end of the second piece of tubing to the antifoam cartridge by pressing until the
connector clicks into place on the inlet valve. Then, push on the end of the syringe to empty
the contents into the cartridge. The syringe has a volume of 100mL and the cartridge a
volume of 280mL, so repeat the process until the antifoam cartridge is full.
8.
On completion, rinse out the syringe and all tubing with water. Screw the cap back onto the
antifoam container and store the tools and container in a cool dry place.
Important Note:
After installation of the antifoam cartridge in the analyzer, it is critical that any
air bubbles be cleared from the antifoam circuit
before
taking any measurements. To do this,
select the
Antifoam system
option from the
Actuator verification
option in the
Maintenance
menu and follow the instructions as described in
. Select 50
injections and stop the process when the system constantly spits a small stream of antifoam at
each stroke denoting the antifoam circuit is free of air bubbles.
Summary of Contents for ORBISPHERE 6110
Page 1: ...DOC024 52 93009 ORBISPHERE Model 6110 Package Analyzer User Manual 08 2013 Edition 5 ...
Page 2: ......
Page 6: ...4 Table of Contents ...
Page 14: ...12 General Information ...
Page 16: ...14 Specifications 2 2 Instrument dimensions Figure 1 Instrument dimensions ...
Page 28: ...26 Installation ...
Page 34: ...32 User Interface and Startup 4 3 Menu structure overview ...
Page 35: ...33 Section 5 Configuration 5 1 Configuration menu overview ...
Page 46: ...44 Configuration ...
Page 56: ...54 Measurement ...
Page 58: ...56 Analysis ...
Page 70: ...68 Spare Parts and Accessories ...
Page 74: ...72 Material Safety Data Sheets MSDS ...
Page 75: ...73 ...