46
Calibration
6.4 Barometric pressure sensor
After selecting this option, the details of the last barometric pressure sensor calibration are
displayed. This shows the values recorded, and information about who performed the last
calibration and when. The due date of the next calibration is also shown (determined by the
parameters entered in
Scheduled operations on page 43
6.5 Pressure sensor
After selecting this option, the details of the last pressure sensor calibration are displayed. This
shows the values recorded, and information about who performed the last calibration and when.
The due date of the next calibration is also shown (determined by the parameters entered in
Scheduled operations on page 43
).
Important Note:
This procedure uses information from the barometric pressure sensor. Be sure
that this parameter is reliable (i.e. a recent barometric pressure sensor calibration has been
performed - see
above).
Barometric pressure sensor
New calibration 1.
A calibration screen is displayed showing the current values measured by the
instrument.
2.
Using a precision certified barometer, measure the barometric pressure in the
location where the instrument is used and compare with the current barometric
pressure value displayed. If the values are the same press
Cancel
, otherwise
enter the new barometric value in the
New barometric value
box and press
Validation
to store the new setting.
3.
On completion a calibration report is created.
Calibration
reports
View the last 10 barometric pressure sensor calibration reports.
Pressure sensor
New calibration 1.
A calibration screen is displayed showing the current values measured by the
instrument.
2.
Using a precision certified barometer, measure the barometric pressure in the
location where the instrument is being used. Provided a barometric pressure
sensor calibration has recently been performed (as described in the
Important
Note
above) this measurement should be the same as the value displayed in the
Reference pressure
box. If this is the case, press
Enter
and go to the next step.
Otherwise, enter the measured barometric pressure value in the
Reference
pressure
box and press
Enter
.
3.
Connect a precision certified pressure gauge (0-7 bar) to the gas out connection
on the rear of the instrument (
No. 4
in
) and measure the
reference pressure. Enter this value in the
Reference pressure
box and press
Enter
.
4.
Finally press the
Validation
button at the bottom left of the screen to accept the
new calibration and complete the process.
5.
On completion a calibration report is created.
Calibration
reports
View the last 10 pressure sensor calibration reports.
Summary of Contents for ORBISPHERE 6110
Page 1: ...DOC024 52 93009 ORBISPHERE Model 6110 Package Analyzer User Manual 08 2013 Edition 5 ...
Page 2: ......
Page 6: ...4 Table of Contents ...
Page 14: ...12 General Information ...
Page 16: ...14 Specifications 2 2 Instrument dimensions Figure 1 Instrument dimensions ...
Page 28: ...26 Installation ...
Page 34: ...32 User Interface and Startup 4 3 Menu structure overview ...
Page 35: ...33 Section 5 Configuration 5 1 Configuration menu overview ...
Page 46: ...44 Configuration ...
Page 56: ...54 Measurement ...
Page 58: ...56 Analysis ...
Page 70: ...68 Spare Parts and Accessories ...
Page 74: ...72 Material Safety Data Sheets MSDS ...
Page 75: ...73 ...