24
Installation
1.
Connect the syringe to the tubing with the quick connector.
2.
Unscrew the cap from the antifoam recharge bottle and put the tubing in, making sure it
goes as near to the bottom of the liquid as possible.
3.
Pull on the syringe to extract the liquid until the syringe is full.
4.
Remove the syringe and tubing from the recharge bottle.
5.
Attach the end of the tubing from the syringe to the other piece of tubing using the quick fit
connector.
6.
Remove any trapped air in the system by applying a small amount of pressure to the end of
the connector with a small screwdriver (or similar) to open the valve. Push slightly on the
end of the syringe until antifoam can be seen dripping out of the end of the connector.
Remove the screwdriver.