38
Configuration
5.4.3 System information
5.5 Security and user management
5.5.1 Security management
Note:
When the instrument is started for the first time, security is disabled by default. To avoid any
unauthorized access, it is highly recommended that each user be entered into the system and given
appropriate access rights as soon as possible. Details are provided in this section.
5.5.2 User management
System information
Display the system information about the currently configured boards. Select
Exit
to exit the screen.
Security management
Configure parameters related to confidentiality. On completion, select
Ok
to confirm.
Access rights
enabled
When checked (recommended), it is required to log on as a registered user to access
the menus. When disabled (default), all menus are access free and there will be no
name recorded against any actions in the log file. To set up valid users, refer to
below.
Auto logoff
When checked, the user is logged out automatically when the set delay for inactivity
(Logout timer) is reached and the instrument goes on standby.
Audit trail
When checked, every action from a user is recorded in an audit file for traceability.
These actions cover the configuration, calibration, and maintenance options. Each
record holds the date and time of the action, the action ID and description, and the
operator name and ID. The audit file is a rolling buffer recording the last 1000 actions.
Logout timer
Enter a maximum time of inactivity (in minutes) for all users. Users are logged out
automatically when the period of inactivity exceeds this value and the Auto logoff
feature is enabled.
Clear audit trail
file
This option deletes the audit file.
Clear result file
This option deletes the package measurement data.
User management
When selected, the list of registered users (up to a maximum of 99) for this instrument are displayed.
Note:
To have the
Delete
and
Edit
options available, select one of the existing users first.
New
To add a new user, all of the following fields must be entered, then select
Ok
to add
the new user.
Name
Enter the user surname (3-15 characters).
First Name
Enter the user first name (3-15 characters)
ID
Enter an alphanumeric ID (1-10 characters).
Password
Enter an alphanumeric password (3-15 characters).
Security level
Select the security level from the drop-down list. For a list of options
available at each level refer to
below.
Edit or Delete
To remove or modify an existing user, select the user in the user management screen
and select
Edit
or
Delete
as appropriate. To save all changes, press
Ok
(you will be
required to confirm the changes) or press the
Exit
button to exit the screen without
making any changes.
Table 2 Security level table
Level
Available options
Operator
Measurement and Analysis
Supervisor
Calibration, Measurement, Analysis and Maintenance
Manager
Configuration, Calibration, Measurement, Analysis and Maintenance
Administrator
Configuration, Calibration, Measurement, Analysis and Maintenance
Summary of Contents for ORBISPHERE 6110
Page 1: ...DOC024 52 93009 ORBISPHERE Model 6110 Package Analyzer User Manual 08 2013 Edition 5 ...
Page 2: ......
Page 6: ...4 Table of Contents ...
Page 14: ...12 General Information ...
Page 16: ...14 Specifications 2 2 Instrument dimensions Figure 1 Instrument dimensions ...
Page 28: ...26 Installation ...
Page 34: ...32 User Interface and Startup 4 3 Menu structure overview ...
Page 35: ...33 Section 5 Configuration 5 1 Configuration menu overview ...
Page 46: ...44 Configuration ...
Page 56: ...54 Measurement ...
Page 58: ...56 Analysis ...
Page 70: ...68 Spare Parts and Accessories ...
Page 74: ...72 Material Safety Data Sheets MSDS ...
Page 75: ...73 ...