Wire Gauge Considerations
Background
Before connecting the output terminals to a load,
the wire gauge of the cables should be considered.
It is essential that the current capacity of the load
cables is adequate. The rating of the cables must
equal or exceed the maximum current rated output
of the instrument.
Recommended
Wire Gauge
Wire Gauge
Nominal Cross
Section
Maximum Current
20
0.5
9
18
0.75
11
18
1
13
16
1.5
18
14
2.5
24
12
4
34
10
6
45
8
10
64
6
16
88
4
25
120
2
32
145
1
50
190
00
70
240
000
95
290
0000
120
340
The maximum temperature rise can only be 60
degrees above the ambient temperature. The
ambient temperature must be less than 30 degrees.