GETTING STARTED
31
Grounding
The output terminals of the ASR-2000 series are isolated with
respect to the protective grounding terminal. The insulation
capacity of the load, the load cables and other connected devices
must be taken into consideration when connected to the protective
ground or when floating.
Grounded Neutral
Output
Basically, grounded return on the neutral output is
allowed for ASR-2000 series and electric shock may
occur if not following the grounding procedure
based on the local electrical safety codes. In some
cases, 0 V is specifically required between ground
and neutral, which can substantially moderate
ground loops, thus keeping sensitive equipment
from effects of ground loops and reducing ground
noise.
Ground & Neutral
Shortcut
Illustration
LINE
NEUTRAL
GROUND
LINE
NEUTRAL
GROUND
CAUTION
Owning to the fact that the neutral has been shortcut
with the ground which is referenced to the chassis
ground, few electric shocks may still take place from
time to time, for which we sincerely ask your
additional attention.