OPERATION
101
Remote Sense
Remote Sense
Operation
Remote sense is used to compensate for the voltage
drop seen across load cables due to resistance
inherent in the load cables. The remote sense
function can compensate a maximum of 5% of the
output voltage and all of output frequency.
1.
Configure the remote sense setting to ON (page
132).
2.
Connect the Neutral terminal of the remote
sense terminal block to the Neutral terminal of
the load.
3.
Connect the Live terminal of the remote sense
terminal block to the Live terminal of the load.
Connection
Example
OUTPUT
terminal
block
Load
cables
Sensing
points
Sensing
cables
N.C. terminals are Not
available in the case.
Load
Note
Do Not connect any wires to the N.C terminals of the
remote sense terminal block.