TEST MODE FUNCTION
161
Time
Sets the duration time of the step.
When the ON Phs=ON, the total
duration of the step is equal to the
Time s ON Phs=ON
duration.
ACV
Sets the voltage of the step. Not
applicable for the Trans 1/2 steps and
the Normal2 step.
ON Phs
Sets the starting phase of the
waveform for the step. Not applicable
for the Trans 1/2 steps.
Fset
Sets the frequency of the step. Not
applicable for the Trans 1/2 steps and
the Normal2 step.
OFF Phs
Sets the off phase of the waveform
after the output has been turned off.
Not applicable for the Trans 1/2 steps.
Wave
Sets the outputting waveform of the
step. Not applicable for the Trans 1/2
steps and the Normal2 step.
Code
Sets the synchronous code including
LL, LH, HL and HH for the duration
of the step.
Repeat
Indicates the number of times the
simulation will be run, from Normal1
to Normal2.
A value of 0 indicates infinite repeats.
The repeat setting is the same for
each step.