TEST MODE FUNCTION
165
10.
Go to the
OFF Phs
setting and set the end phase
of the step.
Not applicable for Trans1 and Trans2.
OFF Phase
Free, Fixed
OFF Phase
0.0 ~ 359.9º
Resolution
0.1º
11.
Go to the
Wave
setting and set the wave of step.
Not applicable for Trans1, Trans2 and Normal2.
Wave
SINE
12.
Go to the
Code
setting and set the synchronous
code of the step.
Code
LL, LH, HL, HH
13.
Lastly, go to the
Repeat
parameter select the
number of times the simulation will repeat the
Normal1-Trans1-Abnormal-Trans2-Normal2
sequence of steps. A value of 0 will set the
number of repetitions to infinite.
Repeat
1 ~ 9999, 0(infinite)