To replace the Charge PCB, pull the emergency cord to
turn the hoist off . Remove the bottom cover .
Note:
It is important to always use a anti-static mat and
wrist band when working inside the hoist .
Disconnect the battery wires as described on page 74 .
Unplug all wires from the PCB and remove from the plastic
housing .
Replace PCB and reconnect wires leaving battery wires
until last .
Note:
It is important to register all PCB exchanges on the
Guldmann traceability web site.