Installation of PCB switch track GH, sparepart
Important
When installation PCB and cables etc . protection against
static electricity must be used .
Un pack the new PC board .
Switch 1 and 7 set to ”ON”
The switches 2-3-4-5-6-8-9-10 set to ”OFF”
Please se electrical diagram on page 155 .
Place the four adhesive circuit board supports in the holes
of the PC board .
Reconnect the motor unit to the PC board according to the
electrical diagram (page 155) .
Remove the protective tape from the circuit board sup-
ports, make sure that the cables are behind the PC board
and place the PC board in the middle of the cover
Re-connect the wires from the transformer according to the
electrical diagram (page 155) .
Fasten the transformer cables with the strain relief – leaving
only the shortest possible cable .