© Guldmann
GB/US
-09/2016 •
#550376_3
41
SERVICE AND
REP
AIR
All PCB’s, plugs and switches must be coated during
service inspection . .
The white circles at the pictures indicate where it is allowed
to spray coating .
Remember to coat both sides at the PCB’s .
NOTICE:
Lifting strap, strap guide and emergency stop
may not be coated .
It is only NON UL hoists which may be coated.
IR PCB must also be coated, but avoid to spay at the
transparent part .
All plugs must be coated .
Charger PCB must also be coated .
Quality inspection:
Be sure that the all PCB’s, plugs and switches are coated
with an evenly distributed layer of coating .
GH3
Coating at GH1/GH3 during annual inspection