Switch track GH
Make sure that the installation arm is installed in between
the legs of the change arm .
Place the two washers and the pin, secure the second set-
screw with Kemlock AS 43 or similar and tighten it .
Re-connect the wires from the transformer according to the
electrical diagram page 155) .
Fasten the transformer cables with the strain relief – leaving
only the shortest possible cable .
Place the cover correct and test that the functions of the
switch track by pulling the turning arm from side to side .
When switching the activation arm shall move from side to
side with a distinct movement . If this is not the case, please
follow the list until correct function is obtained .
i) Control that the change arm and the turning arm are
connected correct .
j) Control that the activation plate and the activation arm
are connected correct . The activation arm must be
placed between the legs of the activation plate .
k) Make sure that the sleeve is placed right above the
motor mounting plate, if not move the sleeve and the
two locking washers to make sure the sleeve is placed
correct and re-install the cover .
l) The sleeve must obtain a slight pressure against the
motor mounting plate . If this is not the case, slightly
bend the activation arm to ensure a slight pressure and
test the function of the switch track .
Re-install the cover and test again .
Summary of Contents for GH1 F
Page 20: ... Guldmann GB US 09 2016 550376_3 22 GENERAL INFORMATION ...
Page 151: ... Guldmann GB US 09 2016 550376_3 153 TECHNICAL DRAWINGS Jumper settings GH3 GH3 Twin ...
Page 158: ... Guldmann GB US 09 2016 550376_3 160 TECHNICAL DRAWINGS ...
Page 164: ... Guldmann GB US 09 2016 550376_3 166 NOTES ...
Page 165: ... Guldmann GB US 09 2016 550376_3 167 NOTES ...
Page 166: ... Guldmann GB US 09 2016 550376_3 168 NOTES ...