Error messages during the reading in of
the
voltage signal of the pH electrode
The calibration may be aborted in the following
cases:
•
If the electrode is faulty or a wrong buffer has
been used.
– The message "fault of electrode/buffer" is
displayed.
•
If the electrode does not reach a stable
measuring signal in 120 seconds.
– The message "calibration time exceeded" is
displayed.
•
If the result of the plausibility testing of the pH
calibration data is upward/downward violation of
the slope range of -50 to -62 mV/pH.
– The message "slope error" is displayed.
•
Downward/upward violation of the
precision/exactness range from -60 to +60 mV.
– The message "error asym. pot." is indicated.
The calibration process will be aborted in all four
cases.
•
Press [OK] to return to the "calibration" menu
automatically.
If the process is aborted, the old calibration date will
not be overwritten!
Manual temperature compensation
With deactivated temperature measurement, the water
temperature for compensation should be entered
manually after calibration. The default value is 25 °C.
1. Determine the water temperature with a
thermometer.
2. Press [Cal] to switch to the calibration function.
Select the line "pH", and press [OK] to switch to
the "pH" calibration menu.
3. Select "manual temp.", and press [OK] to switch
to the corresponding menu.
4. Enter the water temperature.
– The water temperature is shown in °C (Celsius)
or in °F (Fahrenheit), as selected in the "setup"
menu "measuring ranges".
5. Press [OK] to return to the "calibration" menu
automatically.
Cal. cycle
•
Following selection of "cal. cycle" in the
"measured value" menu, a countdown function is
started which triggers the alarm signal "calibrate
sensor" following expiry of a defined time interval
of 1-100 days.
Cal result
•
The result of the current (last) calibration can be
displayed at any time using the "cal result" menu
item.
manual temp.
25 °C
Summary of Contents for Conex DIA-2
Page 2: ...2...
Page 456: ...RU 456 2 Conex DIA 2 1 2 pH Conex DIA 2 PID...
Page 459: ...RU 459 4 Conex DIA 2 GRUNDFOS 5 Conex DIA 2 Cl2 ClO2 O3 H2O2 pH 6 6 1 Conex DIA 2 6 2 GRUNDFOS...
Page 468: ...RU 468 10 1 2 3 Conex DIA 2 4 IP65 IP65...
Page 481: ...RU 481 11 4 1 OK Esc 10 AQC D2 D12 Cal Man Cal Man Cal Man...
Page 488: ...RU 488 11 6 11 6 1 30 1 2 Up Down OK TM03 6734 4506 1 2 2 1 P PI PID...
Page 489: ...RU 489 1 Up Down OK 2 Up Down OK 3 Up Down OK 4 Up Down OK 5 Up Down P PI PID OK...
Page 492: ...RU 492 11 6 4 PI PID 1 33 2 Man 20 20 3 4 OK Esc 5 OK Esc TM03 6715 4506 0 43 pH...
Page 493: ...RU 493 11 7 34 TM03 6737 4506 1 2 1 1 2 2 2 1...
Page 495: ...RU 495 11 8 OK 35 TM03 6738 4506 5 1 10 1 pH 2 4 1 1 1 3 1 4 1 2 0 4 A 10 A 20 A 2 3...
Page 499: ...RU 499 120 pH 50 62 pH 60 60 OK 25 C 1 2 Cal pH OK pH 3 OK 4 C F 5 OK 1 100 25 C...
Page 501: ...RU 501 11 10 1 Man 2 Up Down Man 3 0 4 Man 11 10 1 38 TM03 6721 4506 Man 1 0 100 2 0 100...
Page 503: ...RU 503 11 10 2 39 OK 11 6 3 TM03 6721 4506 Man 1 2 0 100...
Page 504: ...RU 504 12 1 a 2 pH pH 7 a 3 a 4 a b pH c pH d e 5 a 6 pH a b 7 pH a b c 8 a 11 5 8 9 a...
Page 505: ...RU 505 13 14 GRUNDFOS GRUNDFOS 15 24 GRUNDFOS...
Page 608: ...608...
Page 609: ...609...