Automatic water analysis system
GENO-softwatch Komfort
Order no. 145 172 955-inter Edited by: phar-mrie G:\BA-172955-INTER_145_GENO-SOFTWATCH KOMFORT
34
4.2.1 Analysis interval
There are 4 fixed interval periods that can be set by means of the
programme switches S1 and S2. The interval period determines the
frequency with which the analyses are carried out. It is the time
between 2 subsequent starts of an analysis.
Note:
If the IN input (terminal 18/19) is open, no analyses are started.
Upon delivery, this input is bridged.
Note:
The input contact (terminal 18/19) can be switched to a start
function (S10 = ON) via the program switch S10.
If program switch S10 is programmed to ON and the jumper is not
removed at input contact terminal 18/19, analyses are carried out
permanently.
Analysis
interval
S1 S2
Time
[min]
OFF OFF 5
OFF ON 10
ON OFF 20
ON ON 30