Automatic water analysis system
GENO-softwatch Komfort
Order no. 145 172 955-inter Edited by: phar-mrie G:\BA-172955-INTER_145_GENO-SOFTWATCH KOMFORT
24
E Commissioning
The work described below may only be performed by trained experts.
We recommend having Grünbeck's technical service/authorised
service company start up the system.
1 | Start-up
Note:
Make sure that the analysis system has been installed in
accordance with chapter D and that the programme switches have
been programmed according to the respective requirements, refer
to chapter F-4.
Step 1:
Switch on the system
Switch on mains switch
The green LED (L-1) is flashing
Step 2:
Reset the filling volume of the indicator
Simultaneously press keys (T3 and T4)
Note:
Reset filling volume only, if a full bottle was inserted (refer to
chapter F-2).
Step 3:
Filling the measuring chamber with water
Press key (T2) until the measuring chamber is filled up and the
sample is free of air bubbles
Step
4
:
Deaerate the dosing pump
Press key (T3) until indicator continuously flows into the
measuring chamber
Note:
During this time the agitator blade is turning in the measuring
chamber
Step 5:
Start the analysis
Press key (T1) to start the first analysis
Note:
An analysis starts with the flushing of the measuring chamber
L1
T3
T4
T2
T3
T1