Automatic water analysis system
GENO-softwatch Komfort
Order no. 145 172 955-inter Edited by: phar-mrie G:\BA-172955-INTER_145_GENO-SOFTWATCH KOMFORT
33
4.1.2 Setting the delay time
The delay time before sampling (e.g. in case of hot water) can be set
to between 0 seconds and 30 minutes by means of the rotary switch
"delay time" on the right.
Item Time
[min]
Item Time
[min]
Item Time
[min]
Item Time
0 0
4 4 8 8 C 18
1 1 5 5 9 9 D 20
2 2 6 6 A 10 E 25
3 3 7 7 B 12 F 30
The delay time must be selected in a way that definitely no hot water
will be flowing into the measuring chamber at the start of an analysis.
Caution!
If condensate shall be monitored, it is essential that the
sample be cooled down to a temperature of less than 40 °C.
However, in case of longer sampling intervals, it does not make sense
to permanently cool down the condensate. In order to save cooling
water, a cooling water valve must be opened as soon as a sample is
requested. The inlet valve of the analysis device may only be opened
- with delay - if the sample in the cooler has reached a temperature of
less than 40 °C.
The value of the delay time must be determined
on site. It depends on the temperatures of the condensate and the
cooling water as well as on the flow conditions within the system.
In order to determine the proper delay time, you always start out with
a long delay time at the beginning.
4.2 Factory setting of the DIP switches
Table F-1: Settings DIP switch (factory setting)
Refer to
chapter
Flushing time
DIP switch
F-4.2.1
Analysis interval
10 minutes
S1 OFF
S2 ON
F-4.2.2
First value suppression
Yes – ON
S3 ON
F-4.2.3
Functions
Relay 1
Permanent contact
No stop of analysis
S4 ON
S5 OFF
F-4.2.4
Measuring parameters
Total hardness
S6 OFF
S7 OFF
S8 OFF
F-4.2.5
Limit value monitoring
Indication that limit value
has been exceeded
S9 OFF
F-4.2.6
Input function
Flow controller function
S10 OFF
Operating mode
Analysis operations
S11 OFF S12 OFF