Automatic water analysis system
GENO-softwatch Komfort
Order no. 145 172 955-inter Edited by: phar-mrie G:\BA-172955-INTER_145_GENO-SOFTWATCH KOMFORT
46
2.2 Cleaning the measuring chamber
Switch off the system to "0".
Disconnect
the
connection
coupling at the device inlet "IN".
Collect running water in a
container
Pull the stainless steel safety
pins at the measuring chamber
as far as they will go.
Pull the dosing plug, the
inlet plug, the light source
and the outlet plug from the
measuring chamber.
Remove the measuring
chamber from the control box.
Put the measuring chamber
into the cleaning liquid for at
least 10 minutes.
Clean the measuring chamber
mechanically under running
water by means of cleaning
brushes.
Reinstall the measuring
chamber in reverse order in
the system and re-establish
water connection.
Switch on the system again.
Press key: Fill the
measuring chamber with
water. Perform sample
analysis.