Automatic water analysis system
GENO-softwatch Komfort
Order no. 145 172 955-inter Edited by: phar-mrie G:\BA-172955-INTER_145_GENO-SOFTWATCH KOMFORT
47
2.3 Changing the cartridge and hose pump
Note:
To ensure the measuring accuracy of the GENO-softwatch
Komfort, the hose pump cartridge should be replaced after 6 months.
1. Press the locking straps together with your thumb and index finger
and pull the cartridge to the right off the motor shaft.
2. Loosen the bayonet locks of the LUER connections.
3. Pull up the fastening pins of the indicator plug and remove the plugs.
4. Replace the blue O-ring of the indicator plug.
5. Insert the indicator plug into the measuring chamber and secure it
with the fastening pins.
6. If indicator fluid leaks, remove it with a paper towel.
7. Insert a new cartridge in reverse order.
8. Deaerate pump: Switch on the device and press the INDICATOR
key until the hose pump pumps indicator into the measuring
chamber without bubbles.