4 | Electrical wiring
Internal wiring of control unit GENO-OSMO-RO125K
or AVRO 125 TS/TL
The system is completely pre-wired (possibly including options) and
delivered ready to plug in. During commissioning, parameter ECL:1
must be reprogrammed to ECL:0 at code level 113 (NC contact
>>NO contact). This is a protective measure to prevent the system
from being inadvertently switched on after the power cable has been
plugged in, without the system having been vented first.
Jumpers must be plugged in like this
Fig. D-3: Printed circuit board assignment GENO- OSMO RO 125K or AVRO 125 TS/TL
Terminal no.
Terminal Function (core colour)
Note
All protective grounding conductors are connected to the 7-pin protective grounding conductor terminal on the left-
hand mounting rail
X1
3
L
230 V / 50 Hz phase
Mains cable, on-site fuse protection
min. 6 A
2 N Neutral
conductor
1 PE Earth
wire
X2
6
MVW
230 V / 50 Hz phase
Rinse solenoid valve
Common neutral
conductor terminal
5 N Neutral
conductor
7
MVE
230 V / 50 Hz phase
Inlet solenoid valve
10
MVR
230 V / 50 Hz phase
Option:
Solenoid valve forced withdrawal
9 N Neutral
conductor