10-10
10. ADJUSTMENTS USING THE SOFTWARE
10.4.2 Feed distance adjustment
In the following cases, perform this adjustment.
• When the distance is different in the direction of the feed.
• When you replaced the main board.
• When you replaced or removed the CIS sensors.
• When you replaced or removed the feed motor.
• When you replaced or removed the feed rollers.
How to perform the feed distance adjustment
(1) Prepare the CS600/500/IS200 Test Chart.
(2) Measure the distances between the C010 and the C090 points in the CS600/500/IS200 Test Chart with
a glass scale. And then record it.
(3) Run the software. (See page 10-1) This is not necessary when the software is running already.
(4) Select "Adjust Feed Distance" from the "Adjust" menu.
(5) The following dialog box will appear. Correctly insert the CS600/500/IS200 test chart to the scanner, and
then click the "Scan" button.