84
alarm low limit.
n
DEFAULT: select "DEFAULT" to enter "TEMP DEFAULT CONFIG" interface, the user may select "FACTORY
DEFAULT CONFIG" or "USER DEFAULT CONFIG". After selecting, the system will prompt the user to confirm,
then exit.
14.5 TEMP Alarm message
The alarm which triggers by the parameters exceeding the limits, which may activate the recorder to automatically
output the parameters and related measured waveforms when the alarms occur on the condition that the alarm record
switch in the related menu is On.
Tables below describe the possible physiological alarms, technical alarms and prompt messages occurring during TEMP
measurement.
Physiological alarms:
Message
Cause
Alarm Level
T1 HI
Measuring value of channel 1 is above upper alarm limit.
User-selectable
TI LOW
Measuring value of channel 1 is below lower alarm limit.
User-selectable
T2 HI
Measuring value of channel 2 is above upper alarm limit.
User-selectable
T2 LOW
Measuring value of channel 2 is below lower alarm limit.
User-selectable
TD HI
Difference between two channels is larger than upper limit.
User-selectable
Technical alarms:
Alarm Message
Cause
Alarm Level
Remedy
T1 SENSOR OFF
Temperature cable of channel 1
may be disconnected from the
monitor.
LOW
Make sure that the cable is
properly connected.
T2 SENSOR OFF
Temperature cable of channel 2 may
be disconnected from the monitor.
LOW
Make sure that the cable is
properly connected.