138
"NIBP COMM ERR"
The NIBP communication part has
problem.
HIGH
Execute the reset program in the NIBP
menu. If the failure still exists, contact
the manufacturer for repair.
"NIBP LOOSE
CUFF"
The NIBP cuff is not connected
correctly.
LOW
Re-connect the NIBP cuff.
"NIBP AIR LEAK"
The NIBP cuff is not connected
correctly or there are leaks in the
airway.
LOW
Check the connection of each part or
replace with a new cuff. If the failure
still exists, contact the manufacturer for
repair.
"NIBP AIR
PRESSURE ERROR"
Problem happens when measuring
the curve. The system can't
perform measurement, analysis or
calculation.
LOW
Check the connection of each part or
replace with a new cuff. If the failure
still exists, contact the manufacturer for
repair.
"NIBP WEAK
SIGNAL"
Problem happens when measuring
the curve. The system can't
perform measurement, analysis or
calculation.
LOW
Check if the setup of patient type is
correct. Check the connection of each
part or replace with a new cuff. If the
failure
still
exists,
contact
the
manufacturer for repair
"NIBP RANGE
EXCEEDED"
Problem happens when measuring
the curve. The system cannot
perform measurement, analysis or
calculation.
LOW
Check the connection of each part or
replace with a new cuff. If the failure
still exists, contact the manufacturer for
repair.
"NIBP EXCESSIVE
MOTION"
The patient's arm moves.
LOW
Check the connection of each part and
the patient situation. Measure again, if
the failure still exists, contact the
manufacturer for repair.
"NIBP OVER
PRESSURE"
Perhaps folds exist in the airway.
LOW
Check for the smoothness in the airway
and patient situation. Measure again, if
the failure still exists, contact the
manufacturer for repair.
"NIBP SIGNAL
SATURATED"
Problem happens when measuring
the curve. The system cannot
perform measurement, analysis or
calculation.
LOW
Check the connection of each part and
the patient situation. Measure again, if
the failure still exists, contact the
manufacturer for repair.
"NIBP TIME OUT"
Problem happens when measuring
the curve. The system cannot
perform measurement, analysis or
calculation.
LOW
Check the connection of each part and
the patient situation. Measure again, if
the failure still exists, contact the
manufacturer for repair.