136
ectable
connection between the electrode and
lead cable.
"BRADY"
Patient suffers from BRADY.
User-sel
ectable
Check the patient' condition; Check the
connection between the electrode and
lead cable.
"VT>2"
Patient suffers from Arr. of VT>2.
User-sel
ectable
Check the patient' condition; Check the
connection between the electrode and
lead cable.
“MISSED BEATS”
Patient suffers from Arr. of
MISSED BEATS.
User-sel
ectable
Check the patient' condition; Check the
connection between the electrode and
lead cable.
"PNP"
The pacemaker is not paced.
User-sel
ectable
Check the connection of the pacemaker;
check the connection between the
electrode and lead cable; check the
patient' condition.
"PNC"
No pacemaker signal is captured.
User-sel
ectable
Check the patient' condition; Check the
connection between the electrode and
lead cable.
"ECG LEAD OFF or
RESP LEAD OFF"
RL lead of ECG is not connected
correctly or ECG lead is not
connected correctly.
LOW
Check the connection of RL lead or
ECG lead cable.
"V LEAD OFF"
The V lead of ECG is not
connected correctly.
LOW
Check the connection of V lead.
"LL LEAD OFF"
The LL lead of ECG is not
connected correctly.
LOW
Check the connection of LL lead.
"LA LEAD OFF"
The LA lead of ECG is not
connected correctly.
LOW
Check the connection of LA lead.
"RA LEAD OFF"
The RA lead of ECG is not
connected correctly.
LOW
Check the connection of RA lead.
"RESP LEAD OFF"
At least one lead among RA, RL
and LL is not connected correctly.
LOW
Check the connection of RA, RL or LL
lead.
"SpO
2
SENSOR OFF"
SpO
2
sensor is not connected
correctly.
LOW
Check the connection of SpO
2
sensor.
"SpO
2
SEARCH
TIMEOUT"
The pulse signal of the patient is
too small so that the system cannot
perform pulse signal analysis.
HIGH
Check the connection of sensor; Check
the patient' condition.