Tables below describe the possible physiological alarms, technical alarms and prompt messages occurring during NIBP
measurement.
Physiological alarms:
Message
Cause
Alarm Level
SYS HI
NIBP SYS measuring value is above upper alarm limit.
User-selectable
SYS LOW
NIBP SYS measuring value is below lower alarm limit.
User-selectable
DIA HI
NIBP DIA measuring value is above upper alarm limit.
User-selectable
DIA LOW
NIBP DIA measuring value is below lower alarm limit.
User-selectable
MEAN HI
NIBP MAP measuring value is above upper alarm limit.
User-selectable
MEAN LOW
NIBP MAP measuring value is below lower alarm limit.
User-selectable
Technical alarms(display in the prompt area below NIBP value):
Message
Cause
Alarm
Level
Remedy
NIBP SELF TEST
ERROR
Transducer or other
hardware of NIBP module is
incorrect.
HIGH
Stop using measuring function of NIBP
module, notify biomedical engineer or Our
service staff.
NIBP COMM ERR
Communication with NIBP
module is failed.
HIGH
If failure persists, stop using measuring
function of NIBP module, notify
biomedical engineer or Our service staff.
NIBP LOOSE CUFF
Cuff is no properly wrapped
or no cuff exists.
LOW
Properly wrap the cuff.
NIBP AIR LEAK
Cuff, hose or connector is
damaged.
LOW
Check and replace the leaking parts, if
required, notify biomedical engineer or Our
service staff.
NIBP AIR PRESSURE
ERROR
Stable pressure value is not
available. e.g. hoses are
tangled.
LOW
Check if the hoses are tangled, if failure
persists, notify biomedical engineer or Our
service staff.
NIBP WEAK SIGNAL
Cuff is too loose or patient
pulse is too weak.
LOW
Use other methods to measure blood
pressure.
NIBP RANGE
EXCEEDED
Measurement range exceeds
the specified upper limit.
HIGH
Reset NIBP module, if failure persists, stop
using measuring function of NIBP module,
notify biomedical engineer or Our service
staff.
NIBP EXCESSIVE
MOTION
Affected by arm motion,
signal noise is too large or
pulse rate is not regular.
LOW
Make sure that the patient under monitoring
is motionless.
NIBP OVER
Pressure has exceeded the
HIGH
Measure again, if failure persists, stop using