35
NOTE:
l
If certain recording is in process, and another parameter demands alarm recording, it will only be executed
after the earlier recording is finished.
Auto Recording
The monitor starts a recording for 8 seconds according to interval time set in the "TIMING REC TIME" of the
"RECORD SETUP " menu. Refer to the section "RECORD" in system setup for details.
Alarm Recording
n
Parameter Alarm
The monitor records waveforms 4/8/16 seconds before and after the alarm (totally 8, 16 or 32 seconds)
(which can be selected in System Menu).
All parameter values during the alarm will also be recorded.
Two waveforms will be output according to the following rules:
1)
If multiple parameter alarms are switched on and triggered simultaneously, the recorder will print
out those of the highest level. If parameters have the same alarm level, the latest alarm will be printed
out.
2)
If an alarm occurs during the recording of another parameter, it will be printed out after the current recording
is finished.
3)
If many alarms occur at the same time, their waveforms will be stored, and then printed in turn.
n
ST Segment Alarm
The monitor records 2-channel ECG waveforms 4, 8, or 16 seconds prior to and after the ST alarm (totally 8,
16, or 32 seconds) (which can be selected in the menu). All parameter values during the alarm will also be
recorded.
n
Arrhythmia Alarm
The monitor records the waveform 4 seconds prior to and after the alarm (totally 8 seconds). All
measurement results during the alarm will also be recorded.
Freeze Waveform Recording
The monitor prints out the selected waveforms under the FREEZE mode. In this way you can capture the
abnormal waveforms on the screen by freezing and record it.
Trend Graph/Table Recording
The monitor can print out the trend graph and table in the current trend review interface.
Arrhythmia Review Recording
The monitor can print out the arrhythmia alarm event in the current ARR RECALL interface.
Alarm Recall Recording
The monitor can print out the alarm events in the current ALARM RECALL interface.
NIBP Recall Recording
The monitor can print out all the NIBP review events in NIBP RECALL interface.
SD Recall Recording
The monitor can print out the trend data of the case currently review.