If you have questions concerning disposal of the product, please contact our company or representative
institution.
l
When you have questions about the integrity of the external grounding of the monitor and its arrangement,
the internal battery must be used for operation.
l
Electromagnetic fields can affect the performance of the monitor, so other equipment used near the monitor
must meet the appropriate EMC requirements. Mobile phones, X-rays, or MRI devices are possible sources
of interference because they could emit high-intensity electromagnetic radiation.
l
Before turning on the power to the device, make sure that the supply voltage and frequency match the
device's label or the requirements specified in this manual.
l
When the battery is about to exceed its service life, remove the battery immediately from the monitor.
l
To ensure patient safety, please use the accessories specified in this manual.
NOTE:
l
Install the equipment in a location that is easy to observe, operate and maintain.
l
If the monitor gets damp accidentally, or the liquid is dumped on the equipment or accessories, especially if
the liquid is likely to enter the monitor, please contact the service personnel in time.
l
The software is developed in accordance with IEC62304. The possibility of risks caused by program error
has been minimized.
l
The pictures and interfaces in this manual are for reference only, please in kind prevail.
1.2 Precautionary measures
l
In order to avoid the accumulation of electrostatic charge, it is recommended to store, maintain and use the
equipment at a relative humidity of 30% or more. The floor should be covered with ESD dissipated carpets or
similar materials. In the use of the components, non-synthetic clothing should be wore.
l
In order to prevent electrostatic discharging to the ESD-sensitive parts of the device, the personnel should contact
the metal frame of the components or the large metal objects near the device. When using the device, especially
when it is possible to contact the ESD-sensitive parts of the device, the operator should wear a grounded bracelet
designed for ESD-sensitive devices. For more information on proper use, please refer to the instructions provided
with the bracelet.
ESD Precautionary procedure training
l
All potential users are advised to understand the ESD warning symbols and receive training on ESD precautions.
l
The most basic content of the ESD precautionary procedure training should include an introduction to electrostatic
charge physics, voltage level in the conventional case, and damage to the electronic components when the operator
with electrostatic charge contacts them. In addition, the methods for preventing electrostatic buildup, and the
manner and reasons for the release of human body static electricity to the ground or equipment frame or the use of
a bracelet to connect the human body to the equipment or the ground before establishing the connection should be
described.