56
10.4 ECG Screen Hot Keys
The following figure is an interface of 5-lead monitoring, only for reference.
ECG Hot Key
Errore. L'origine riferimento non è stata trovata.
Leads of channel 1:
1)
The selectable leads are I, II, III, aVR, aVL, aVF, V.
2)
When the ECG is 5-lead, the selectable leads are: I, II, III, aVR, aVL, aVF; V. When ECG is 3-lead, the
selectable leads are: I, II, III.
3)
Leads on the ECG waveform must not use the same name. Otherwise, the system will automatically change
the ECG waveform name that has been used to another one.
Errore. L'origine riferimento non è stata trovata.
Waveform gain of channel 1: to adjust the amplitude of ECG
waveforms
Select gain value for each channel from
×
0.25,
×
0.5,
×
1,
×
2 and
×
4. A 1mV scale displays on each ECG
channel's one side. The height of the scale is directly proportional to the waveform amplitude.
Errore. L'origine riferimento non è stata trovata.
Filter method: to display clearer and more detailed waveform
There are three filter modes for selection. In DIAGNOSTIC mode, the ECG waveform is displayed
without filter. In MONITOR mode, the artifact that may cause false alarm is filtered. And the
SURGERY mode could reduce artifacts and interferences from electrosurgery equipment. The filter
mode is applicable for both channels, and it is displayed at the top of screen.
Errore. L'origine riferimento non è stata trovata.
Leads of channel 2: refer to
①
for detailed information.
Errore. L'origine riferimento non è stata trovata.
Waveform gain of channel 2: refer to
②
for detailed
information.
WARNING
②
③
④
⑤
①