12.6 SpO
2
SETUP
Turn the knob to move the cursor onto the SpO
2
hot key in the Parameter area, push the knob to "SpO
2
SETUP" menu.
n
ALM REC: pick "ON", the system will output alarm information when SpO
2
alarm occurs.
n
SWEEP: 12.5mm/s, 25.0 mm/s
n
SpO
2
alarm setting
Set the following contents after entering the initial password "70808":
u
SpO
2
ALM: pick "ON", the system will give alarm prompt and store alarm information when SpO
2
alarm
occurs; pick "OFF", the system will not give alarm and instead display a
beside “SpO
2
”.
u
ALM LEV: set the alarm level, selectable from HI, MED and LO. HIGH represents the most serious case.
u
SpO
2
ALM HI and SpO
2
ALM LO: SpO
2
alarm is activated when the result exceeds set SpO
2
ALM HI value
or falls below SpO
2
ALM LO value.
u
PR ALM: pick "ON", the system will give alarm prompt and store alarm information when PR alarm
occurs.
u
PR ALM HI: PR alarm is activated when the pulse rate exceeds set PR ALM HI value.
u
PR ALM LO: PR alarm is activated when the PR falls below PR ALM LO value.
To further detect alarms for individual measurement parameters, perform a measurement check on yourself or by using
the simulator, adjust the alarm limits setting and check if the correct alarm response is triggered.
WARNING
l
Set the upper limit of SpO
2
alarm to completely equal to off-state upper limit alarm. High-oxygen level will
cause fibrous fibrosis for preterm infants. Therefore, the upper limit of the SpO
2
alarm must be carefully
chosen according to accepted clinical practice.
NOTE:
l
The upper and lower limit of SpO
2
alarm will be displayed continuously in the SpO
2
parameter area.
n
DEFAULT: select it to "SpO
2
DEFAULT CONFIG" menu, in which you can select "FACTORY DEFAULT
CONFIG" or "USER DEFAULT CONFIG". After selecting one item and exiting the interface, the system will pop
up the dialog box asking for your confirmation.
12.7 SpO
2
Alarm message
NOTE:
l
There is no alarm delay for SpO
2
.
SpO
2
alarm information
When the alarm switches are set to "ON" in relevant menus, the physiological alarms caused by the parameter
exceeding the alarm limit may possibly trigger the recorder to automatically output the alarm parameter value and
corresponding waveforms.
Tables below describe the possible physiological alarms, technical alarms and prompt messages occurring during SpO
2