Working with Charger Connected to the Pipette

PIPETMAN M Connected can also be used for pipetting while charging. 
1. 

Plug the power supply into your pipette.

2.  Press the push button to start your pipette.
PIPETMAN M Connected is ready to pipette.

Low Battery Warning

PIPETMAN M Connected has a low battery alert. As the battery runs 

down, the low battery message appears. The warning is replaced by a 

blinking symbol of an empty battery. If it is not recharged, the pipette will switch off 

after a while. Settings will not be lost.
If the battery is completely discharged (= black screen, no operation) connect the 

pipette to the charger for at least five minutes, and then a “batt. too low” message 

will appear. The pipette will display the last settings used a few minutes later and will 

continue charging.

Changing the Battery

If the following symptoms occur, please contact your Gilson Service Center:

 

PIPETMAN M Connected battery won’t charge or will not operate even if 

connected to the charger at least for ten minutes.

 

PIPETMAN M Connected battery provides you very short cycle time or 

permanently needs to be used in connected mode.

NOTE

You can also recharge PiPETMAN M Connected by USB with your PC; 

however it will take much longer than by power supply.

NOTICE

Use only the original power transformer supplied by Gilson. Use of an 

incompatible power supply can damage your PiPETMAN M Connected!

do not operate PiPETMAN while connected to a PC.

CAUTION

PiPETMAN M Connected uses a Lithium-ion battery. Operation on the 

battery is done at user’s risk only. Opening the pipette voids the warranty.

Alarms

 

Low battery warning 

 

Service is overdue warning (refer to Chapter 15  

MAiNTENANCE

 on page 22)

 

Service alert can be set 

“On” 

or 

“Off”

Chapter 13 

POWEr MANAGEMENT

PIPETMAN M Connected battery has been 

designed to ensure 900 pipetting cycles  

(750 for PIPETMAN M Connected Multichannel 

200 µL and 300 µL) at maximum speed (up 

to nine 96–well plates) between two charge 

cycles. For service continuity when needing 

more capacity, your PIPETMAN M Connected 

can also be used while charging with the same 

performance as a fully charged pipette.

Battery Charging

PIPETMAN M Connected charges 80% of its full 

battery capacity in less than an hour, and it takes 

three hours to fully charge the battery.

Charge your pipette by using the power supply: 
1. 

Connect the AC adapter to a suitable  

AC power supply.

2.  Plug the adapter into the USB port  

of the pipette.

The pipette starts charging.
Charge your pipette by using stand adapter (refer to 

Chapter 2

 

PArTS ChECkLiST ANd ACCESSOriES

 

on page 5): 
1. 

First, place the stand adapter on your Gilson 

SINGLE Pipette Holder or CARROUSEL.

2.  Remove the plug. Now, place your pipette on the 

Stand Adapter and plug the power transformer 

into the stand adapter.

The pipette starts charging.
Battery charging of PIPETMAN M Connected using 

the POWER CARROUSEL: 
1. 

First, install the POWER CARROUSEL as 

described above.

2.  Remove the plug. Now, place your pipette on the 

charging position, please ensure  

that the pipette is properly fitted to the contacts 

on the top of the carrousel (a proper fitting will 

trigger a beep sound).

The pipette starts charging.

Stand adapter

Plug

USB port

Figure 11

PIPETMAN® M Connected power supply

Figure 12

POWER CARROUSEL™ for  

PIPETMAN® M Connected 

Summary of Contents for F81040

Page 1: ...User s Guide EN...

Page 2: ...D 18 Tip Ejector Extension for Use with P10M 18 Personal Label Name and or Application Tag 19 GOOD LABORATORY PRACTICE FEATURES 19 General 19 Cycle Counters 19 Alarms 20 POWER MANAGEMENT 20 Battery Ch...

Page 3: ...fatigue and increase pipetting efficiency PIPETMAN M Connected is a unique pipette with only two buttons to operate the pipette and reach all menu options PIPETMAN M Connected features the classic PIP...

Page 4: ...9 10 11 12 13 14 15 16 17 18 19 Upper part Push button and volume adjustment knob Menu button direct access to all menu functions Tip ejector button Display Elastomer plug to cover the battery chargi...

Page 5: ...the tip by pressing the ejection button Your PIPETMAN M Connected is ready for the next pipetting cycle while maintaining the volume and the last settings Continue pipetting as you would with your mec...

Page 6: ...g protocol first change the pipetting speed of your pipette refer to Chapter 8 PIPETTING SPEED CONTROL on page 15 Aspiration can be done by using the standard speed by default speed 6 Dispensing shoul...

Page 7: ...nables personalized pipetting protocol creation on a PC or tablet for transfer to PIPETMAN M Connected through PIPETMAN M Connected Software an easy to use software To use this mode on a PC the softwa...

Page 8: ...pending on your application You can change aspiration speed and dispensing speed independently from very slow to very fast speed 1 speed 6 Your speed selection is memorized for each pipetting mode unt...

Page 9: ...4 5 P10mLM 5 7 4 5 P8x10M P12x10M 1 1 P8x20M P12x20M 2 3 1 P8x100M P12x100M 2 4 1 P8x200M P12x200M 2 3 1 P8x300M P12x300M 2 3 1 P8x1200M P12x1200M 2 4 2 3 Table 1 Immersion depth and wait time Mold an...

Page 10: ...surface of the liquid refer to Immersion Depth and Wait Time Table 1 on page 17 Change the tip before aspirating a different liquid sample or reagent Change the tip if a droplet remains at the end of...

Page 11: ...thium Ion battery Operation on the battery is done at user s risk only Opening the pipette voids the warranty Alarms Low battery warning Service is overdue warning refer to Chapter 15 MAINTENANCE on p...

Page 12: ...mber of readjustments carried out by this pipette NOTE Changes of service settings will reset all the service info to zero Chapter 14 CONFIGURATION Enter the menu using the menu button You will find n...

Page 13: ...tion Enter the sub menu in the same way and select the ADJUSTMENT menu The adjustment menu allows you to access the following items Standard readjustment Reset settings Standard Readjustment This menu...

Page 14: ...Chapter 21 REPLACEMENT PARTS on page 34 To access the O ring remove the tip ejector and unscrew the lower part of the tip holder You should now be able to remove the O ring from the piston Sometimes t...

Page 15: ...er 16 CLEANING AND DECONTAMINATION PIPETMAN M Connected is designed so that the parts normally in contact with contaminants can easily be cleaned and decontaminated As part of your quality system you...

Page 16: ...nents with distilled or sterile water 4 Leave the parts to dry by evaporation or wipe them with a clean lint free tissue or soft cloth 5 Lubricate the piston and reassemble the pipette according to th...

Page 17: ...e used with the recommended PIPETMAN DIAMOND Tips Under these conditions Gilson volumetric specifications in standard pipetting PIPET mode are guaranteed with a performance exceeding ISO 8655 2 recomm...

Page 18: ...22 F807022 F807022 F807022 F807022 F807022 F807022 Battery window F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005 Ejector window F807006 F807006 F807006 F807006 F807006 F807006 F807006...

Page 19: ...ture and pressure for distilled water TEMPERATURE AIR PRESSURE HPA C 800 853 907 960 1013 1067 15 1 0018 1 0018 1 0019 1 0019 1 0020 1 0020 15 5 1 0018 1 0019 1 0019 1 0020 1 0020 1 0021 16 1 0019 1 0...

Page 20: ...R IMPLIED INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL PUNITIVE INDIRECT OR IN...

Page 21: ...gilson com contactus LT801594 G 2021 Gilson Inc All rights reserved...

Reviews: