PIPETMAN® M CONNECTED 

|

 

USER’S GUIDE

PIPETTING MODES

 

PIPETMAN® M CONNECTED

2.  Set the

 MiX

 volume: click the push button and set the volume by turning the push 

button. Confirm your settings with a simple click. This is the volume that will be 

aspirated and dispensed repeatedly. 

MiX” 

as long as you press the push button.  

3.  Fit a tip suitable for the model of PIPETMAN M Connected that you are using 

(preferably use PIPETMAN DIAMOND Tips for assurance of accurate and precise 

results; refer to Chapter 9 

PiPETMAN diAMONd TiPS

 on page 15).

Mix

Press the push button; as long as you keep your thumb pressing the button, the pipette 

continues mixing. Release the push button to complete the current mixing cycle. 

Purge

Press the push button. 

PUrGE 

appears on the screen. Click again to purge and reset the 

piston.

NOTICE

Always fit a tip before using any pipette!

Figure 4

MIX Mode

REVERSE Mode

The REVERSE Mode is reverse pipetting with a manual pipette. During 

aspiration, additional liquid is added. After delivery, the excess volume 

remains in the tip and is discarded. 
The REVERSE Mode is ideal for pipetting viscous liquids, liquids with high vapor 

pressure, or those that tend to foam. Your PIPETMAN M Connected aspirates a 

selected volume and an excess. This excess compensates for the liquid that remains 

as a film inside the tip during dispensing. For example: protein extraction, cell lysis, 

plasmid isolation, cell culture, buffer preparation and, many others.

Electrophoresis Gels Loading Protocol

Electrophoresis gels loading protocol: first change the pipetting speed of your 

pipette(refer to Chapter 8

 

PiPETTiNG SPEEd CONTrOL

 

on page 15). Aspiration can be 

done by using the standard speed (by default speed 6). Dispensing should be done very 

slowly and carefully, to prevent swirling and spilling of the samples, so select the lowest 

speed 1.  

1. 

Press the menu button to access system menu, where you can choose between 

different pipette modes. Select 

“rEVErSE” Mode

 by turning the push button. 

To confirm your selection click, the push button.

Set the Volume

1. 

Press the push button, set the volume by turning the push button and press again 

to confirm your settings. 

2.  Fit a tip suitable for the model of PIPETMAN M Connected that you are using 

(preferably use PIPETMAN DIAMOND Tips for assurance of accurate and precise 
results; refer to Chapter 9 

PiPETMAN diAMONd TiPS

 on page 15).

Aspirate

Press the push button to aspirate the selected 

volume. An amount of liquid equal to the amount of 

purged air is added. The volume aspirated will be a 

little bit more than the set volume.

dispense

Press the push button to dispense the volume,  

the additional amount of liquid remains in the tip.

Purge

Press the push button. 

PUrGE

 appears on the 

screen. Press again to purge and reset the piston. 

Keep your thumb pressing the push button until 

the end of the purge. The piston reset takes place 

after you release the push button again.  

This feature enables you to remove the tip from 

the liquid, without aspirating anything.

Eject tip

Press the ejection button.

NOTICE

Always fit a tip before using any pipette.

0

Purge 

START button
Optional STOP

Start Purge

V (+ extra vol)*

Click to Purge

Figure 5

REVERSE Mode

NOTICE

if you purge without pulling out the tip, then air bubbles could get into 

your gel!

Summary of Contents for F81040

Page 1: ...User s Guide EN...

Page 2: ...D 18 Tip Ejector Extension for Use with P10M 18 Personal Label Name and or Application Tag 19 GOOD LABORATORY PRACTICE FEATURES 19 General 19 Cycle Counters 19 Alarms 20 POWER MANAGEMENT 20 Battery Ch...

Page 3: ...fatigue and increase pipetting efficiency PIPETMAN M Connected is a unique pipette with only two buttons to operate the pipette and reach all menu options PIPETMAN M Connected features the classic PIP...

Page 4: ...9 10 11 12 13 14 15 16 17 18 19 Upper part Push button and volume adjustment knob Menu button direct access to all menu functions Tip ejector button Display Elastomer plug to cover the battery chargi...

Page 5: ...the tip by pressing the ejection button Your PIPETMAN M Connected is ready for the next pipetting cycle while maintaining the volume and the last settings Continue pipetting as you would with your mec...

Page 6: ...g protocol first change the pipetting speed of your pipette refer to Chapter 8 PIPETTING SPEED CONTROL on page 15 Aspiration can be done by using the standard speed by default speed 6 Dispensing shoul...

Page 7: ...nables personalized pipetting protocol creation on a PC or tablet for transfer to PIPETMAN M Connected through PIPETMAN M Connected Software an easy to use software To use this mode on a PC the softwa...

Page 8: ...pending on your application You can change aspiration speed and dispensing speed independently from very slow to very fast speed 1 speed 6 Your speed selection is memorized for each pipetting mode unt...

Page 9: ...4 5 P10mLM 5 7 4 5 P8x10M P12x10M 1 1 P8x20M P12x20M 2 3 1 P8x100M P12x100M 2 4 1 P8x200M P12x200M 2 3 1 P8x300M P12x300M 2 3 1 P8x1200M P12x1200M 2 4 2 3 Table 1 Immersion depth and wait time Mold an...

Page 10: ...surface of the liquid refer to Immersion Depth and Wait Time Table 1 on page 17 Change the tip before aspirating a different liquid sample or reagent Change the tip if a droplet remains at the end of...

Page 11: ...thium Ion battery Operation on the battery is done at user s risk only Opening the pipette voids the warranty Alarms Low battery warning Service is overdue warning refer to Chapter 15 MAINTENANCE on p...

Page 12: ...mber of readjustments carried out by this pipette NOTE Changes of service settings will reset all the service info to zero Chapter 14 CONFIGURATION Enter the menu using the menu button You will find n...

Page 13: ...tion Enter the sub menu in the same way and select the ADJUSTMENT menu The adjustment menu allows you to access the following items Standard readjustment Reset settings Standard Readjustment This menu...

Page 14: ...Chapter 21 REPLACEMENT PARTS on page 34 To access the O ring remove the tip ejector and unscrew the lower part of the tip holder You should now be able to remove the O ring from the piston Sometimes t...

Page 15: ...er 16 CLEANING AND DECONTAMINATION PIPETMAN M Connected is designed so that the parts normally in contact with contaminants can easily be cleaned and decontaminated As part of your quality system you...

Page 16: ...nents with distilled or sterile water 4 Leave the parts to dry by evaporation or wipe them with a clean lint free tissue or soft cloth 5 Lubricate the piston and reassemble the pipette according to th...

Page 17: ...e used with the recommended PIPETMAN DIAMOND Tips Under these conditions Gilson volumetric specifications in standard pipetting PIPET mode are guaranteed with a performance exceeding ISO 8655 2 recomm...

Page 18: ...22 F807022 F807022 F807022 F807022 F807022 F807022 Battery window F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005 Ejector window F807006 F807006 F807006 F807006 F807006 F807006 F807006...

Page 19: ...ture and pressure for distilled water TEMPERATURE AIR PRESSURE HPA C 800 853 907 960 1013 1067 15 1 0018 1 0018 1 0019 1 0019 1 0020 1 0020 15 5 1 0018 1 0019 1 0019 1 0020 1 0020 1 0021 16 1 0019 1 0...

Page 20: ...R IMPLIED INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL PUNITIVE INDIRECT OR IN...

Page 21: ...gilson com contactus LT801594 G 2021 Gilson Inc All rights reserved...

Reviews: