PIPETMAN® M CONNECTED
|
USER’S GUIDE
LEAK TEST
|
PIPETMAN® M CONNECTED
Chapter 18
TrOUBLEShOOTiNG
In case of malfunction, first reset the pipette by pressing on the push button and the
menu button simultaneously for at least ten seconds.
If the problem persists, you may consult the following table that identifies potential
problems and their solutions.
WARNING
Before returning any pipette to your local Gilson service center, ensure that
it is completely free of chemical, biological, or radioactive contamination.
Please use the included safety bag.
Chapter 19
SPECiFiCATiONS
PIPETMAN M Connected is a high quality pipette that offers excellent accuracy and
precision; it is fully compliant with ISO 8655 and is CE marked.
The figures given in the
Gilson Maximum Permissible Errors
Table 2 on page 32 were
obtained using “PIPETMAN DIAMOND Tips”. These figures are only guaranteed when
using genuine PIPETMAN DIAMOND Tips.
Each pipette is inspected and validated by qualified technicians according to the Gilson
quality system. Gilson declares that its manufactured pipettes comply with the
requirements of the ISO 8655 standard, by type testing.
Non-immersible Parts
1.
Wipe the handle of the pipette with a soft cloth or lint-free tissue covered with
the chosen decontaminant.
2. Wipe the handle of the pipette with a soft cloth or lint-free tissue covered with
distilled water.
immersible Parts
Only the following components can be immersed in a decontaminant solution:
●
Single channel models
a
tip ejector, tip holder (both parts), connecting nut,
piston (including holder), and guide.
●
Multichannel models
a
tip ejector, ejector locks, and ejector spacer.
Please note that although the lower part of PIPETMAN M Connected multichannel can
withstand a few number of autoclaving cycles, we do not recommend autoclaving it.
1.
Disassemble the pipette as described in Chapter 15
on page 22).
2. Immerse the components in the decontaminant solution or wipe them according
the intructions given by the manufacturer or supplier of the decontaminant.
3. Rinse the individual components with distilled or sterile water.
4. Leave the parts to dry by evaporation or wipe them with a clean, lint-free tissue
or soft cloth.
5. Lubricate the piston and reassemble the pipette according to the instructions
given in this chapter.
Chapter 17
LEAk TEST
This test may be performed at any time to check that the pipette does not leak,
especially after performing a maintenance or decontamination procedure. If a pipette
fails this test, you should replace the faulty part (e.g., O-ring, tip holder) and repeat this
test, after making sure that the pipette is correctly reassembled.
1.
Fit a PIPETMAN DIAMOND Tip.
2. Set the pipette to the nominal volume.
3. Aspirate the nominal volume from a beaker of distilled water.
4. Hold the pipette in the vertical position and wait for 20 seconds.
If a water droplet appears at the end of the tip, there is a leak (refer to Chapter 18
on page 31)
If you see no droplet, re-immerse the tip below the surface of water.
The water level inside the tip should remain constant; if the level goes down then there is a
leak (refer to Chapter 18
on page 31)
For multichannel models, check if the water level between each tip is the same.
PROBLEM
POSSIBLE CAUSE
PAGE
Pipette is leaking
sample
Worn O-ring
Pipette won’t
aspirate
Unscrewed lower part
of tip holder
Damaged or corroded piston
Damaged tip holder
Improper repair or assembly
Connecting nut is loose.
Software needs to be reset
Noisy operation
Piston needs lubricating
Pipette is
inaccurate
Improper repair or assembly
Unscrewed lower part
of tip holder
Pipette is out of adjustment
Connecting nut is loose
PROBLEM
POSSIBLE CAUSE
PAGE
Pipette is not precise
Unscrewed lower part of tip holder
26
Incorrect operator technique
17
Worn O-ring
26
Connecting nut is loose
Damaged or corroded piston
26
Damaged tip holder
25
Tips fall or don’t fit
Low quality tips
15–16
Damaged tip holder
25
Damaged tip ejector
25–27
Ejector spacer is damaged
27
Tip ejector is loose
25–27
Dirty tip holder
25
No OLED display
Pipette is in sleep mode
Battery needs recharging
15
Software needs to be reset.
No operation possible
Battery needs recharging.
15
Firmware needs to be reset
Calibration impossible
Firmware needs to be reset