PIPETMAN® M CONNECTED 

|

 

USER’S GUIDE

CONFIGURATION

 

PIPETMAN® M CONNECTED

Service Information

PIPETMAN M Connected provides you with all required service information. It will help 

you establish an easy diagnosis and plan any service operation with your accredited 

Gilson service center. 
You will find in the 

Service

 menu the following sub menus:

 

Disassembly

 

Service info

 

Service settings

 

Pipette info

The navigation of the menu is always done in the same way. Make your selection by 

rotating the push button, enter and confirm with a simple click.

 

disassembly:

 

 

The multichannel models should not be disassembled: only the push button 

and the tip ejector can be replaced

 

Only the lower part of the single models can be disassembled, the push button, 

the connecting nut, and the tip ejector can be replaced

 

This option allows you to disassemble the lower part safely (protection of the 

piston and the actuator). If you select

 diSASSEMBLY

,

 the following screen 

appears: 

“DISASSEMBLY – Click to exit”

 

Service info

 provides you with maintenance information; you cannot modify any of 

the items: 

 

Number of days since last service

 

Number of pipetting cycles since last service

 

Service settings

 gives you the option to organize your maintenance, you can set 

following service points: 

 

Period (days until next service)

 

Pipetting cycles (cyles until next service)

 

Date

 

On/Off

 

Pipette info 

provides information about your pipette, you cannot modify any of 

the items: 

 

Version of firmware

 

Serial number

 

Total number of cycles (one cycle: up and down of the piston)

 

Number of readjustments carried out by this pipette

NOTE

Changes of service settings will reset all the service info to zero.

Chapter 14 

CONFiGUrATiON

Enter the menu using the menu button. You will find not 

only the pipetting programs and the speed control, but also 

the menu 

“CONFIGURATION”,

 including 

“SERVICE” 

and 

“ADJUSTMENT”

 (refer to Chapter 1

MAiNTENANCE 

below).

The navigation of the menus is always done in the same way 

—make your selection by rotating the push button and then 

enter and confirm with a click.
The 

Configuration

 menu allows you to set up the following items:

 

Volume limit:

 Here you can fix the maximal volume of the 

pipette (by default = nominal volume). For example, you 

have a P200M pipette, but you would like to use it with 

your PIPETMAN Diamond Filter Tips DF100, you can set 

the maximal volume at 100 µL and you won’t risk contamination of your pipette. 

 

Beeper: 

You can switch the tone of your pipette on or off.

 

Contrast: 

You can choose a value of 1 to 6 to set the contrast  

of the screen.

 

Service:

 refer to Chapter 15 

MAiNTENANCE 

below. 

 

Adjustment: 

refer to Chapter 15 

MAiNTENANCE

 below. 

Menu button

Push button

Chapter 15 

MAiNTENANCE

PIPETMAN M Connected requires very little maintenance. However, to ensure pipette 

accuracy, precision and robustness please proceed periodically with a two-minute 

inspection as recommended by Gilson. 
Your PIPETMAN M Connected allows you to:

 

Get service information 

 

Readjust the pipette to user settings 

 

Return to factory settings 

 

Replace spare parts (for detailed information refer to Chapter 1

MAiNTENANCE

 on 

page 22 an

Safety Precautions and Limitation of Use 

on page 33)

 

Prepare the pipette for cleaning or autoclaving by “disassembly” of the parts 

specified (refer to Chapter 1

CLEANiNG ANd dECONTAMiNATiON

 on page 28)

Summary of Contents for F81040

Page 1: ...User s Guide EN...

Page 2: ...D 18 Tip Ejector Extension for Use with P10M 18 Personal Label Name and or Application Tag 19 GOOD LABORATORY PRACTICE FEATURES 19 General 19 Cycle Counters 19 Alarms 20 POWER MANAGEMENT 20 Battery Ch...

Page 3: ...fatigue and increase pipetting efficiency PIPETMAN M Connected is a unique pipette with only two buttons to operate the pipette and reach all menu options PIPETMAN M Connected features the classic PIP...

Page 4: ...9 10 11 12 13 14 15 16 17 18 19 Upper part Push button and volume adjustment knob Menu button direct access to all menu functions Tip ejector button Display Elastomer plug to cover the battery chargi...

Page 5: ...the tip by pressing the ejection button Your PIPETMAN M Connected is ready for the next pipetting cycle while maintaining the volume and the last settings Continue pipetting as you would with your mec...

Page 6: ...g protocol first change the pipetting speed of your pipette refer to Chapter 8 PIPETTING SPEED CONTROL on page 15 Aspiration can be done by using the standard speed by default speed 6 Dispensing shoul...

Page 7: ...nables personalized pipetting protocol creation on a PC or tablet for transfer to PIPETMAN M Connected through PIPETMAN M Connected Software an easy to use software To use this mode on a PC the softwa...

Page 8: ...pending on your application You can change aspiration speed and dispensing speed independently from very slow to very fast speed 1 speed 6 Your speed selection is memorized for each pipetting mode unt...

Page 9: ...4 5 P10mLM 5 7 4 5 P8x10M P12x10M 1 1 P8x20M P12x20M 2 3 1 P8x100M P12x100M 2 4 1 P8x200M P12x200M 2 3 1 P8x300M P12x300M 2 3 1 P8x1200M P12x1200M 2 4 2 3 Table 1 Immersion depth and wait time Mold an...

Page 10: ...surface of the liquid refer to Immersion Depth and Wait Time Table 1 on page 17 Change the tip before aspirating a different liquid sample or reagent Change the tip if a droplet remains at the end of...

Page 11: ...thium Ion battery Operation on the battery is done at user s risk only Opening the pipette voids the warranty Alarms Low battery warning Service is overdue warning refer to Chapter 15 MAINTENANCE on p...

Page 12: ...mber of readjustments carried out by this pipette NOTE Changes of service settings will reset all the service info to zero Chapter 14 CONFIGURATION Enter the menu using the menu button You will find n...

Page 13: ...tion Enter the sub menu in the same way and select the ADJUSTMENT menu The adjustment menu allows you to access the following items Standard readjustment Reset settings Standard Readjustment This menu...

Page 14: ...Chapter 21 REPLACEMENT PARTS on page 34 To access the O ring remove the tip ejector and unscrew the lower part of the tip holder You should now be able to remove the O ring from the piston Sometimes t...

Page 15: ...er 16 CLEANING AND DECONTAMINATION PIPETMAN M Connected is designed so that the parts normally in contact with contaminants can easily be cleaned and decontaminated As part of your quality system you...

Page 16: ...nents with distilled or sterile water 4 Leave the parts to dry by evaporation or wipe them with a clean lint free tissue or soft cloth 5 Lubricate the piston and reassemble the pipette according to th...

Page 17: ...e used with the recommended PIPETMAN DIAMOND Tips Under these conditions Gilson volumetric specifications in standard pipetting PIPET mode are guaranteed with a performance exceeding ISO 8655 2 recomm...

Page 18: ...22 F807022 F807022 F807022 F807022 F807022 F807022 Battery window F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005 Ejector window F807006 F807006 F807006 F807006 F807006 F807006 F807006...

Page 19: ...ture and pressure for distilled water TEMPERATURE AIR PRESSURE HPA C 800 853 907 960 1013 1067 15 1 0018 1 0018 1 0019 1 0019 1 0020 1 0020 15 5 1 0018 1 0019 1 0019 1 0020 1 0020 1 0021 16 1 0019 1 0...

Page 20: ...R IMPLIED INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL PUNITIVE INDIRECT OR IN...

Page 21: ...gilson com contactus LT801594 G 2021 Gilson Inc All rights reserved...

Reviews: