PIPETMAN® M CONNECTED 

|

 

USER’S GUIDE

BENEFIT FROM THE BLUETOOTH® CONNECTION

 

PIPETMAN® M CONNECTED

Connection with the PIPETMAN M Connected Software

When it is necessary to have the pipette connected, the software will prompt you. 

If the pipette has been registered previously, it is only necessary to activate the 

Bluetooth on the pipette to make the connection; however, if this is the first time the 

pipette is being used in the software, it is necessary to register the pipette.
1. 

From the main menu, select 

Other administration

 and then 

Pipette management

2.  Select 

 next to 

Registered pipettes

.

3.  Connect the pipette. 
If connecting via Bluetooth:

 

On the tablet, select 

. The software searches for pipettes with Bluetooth 

activated (indicator lights flashing). 

1. 

On the PC, select 

Connect via BLE

. The software searches for pipettes with 

Bluetooth activated (indicator lights flashing).

NOTE

Many devices may be detected, but only one can be connected.

If connecting via USB:

 

Use a USB cable to connect the pipette to the PC.

1. 

Select 

Connect via USB

. The connected pipette appears in the 

Devices 

list

2.  Select the pipette in the 

Devices

 list.

3.  Select 

Next

.

4.  When prompted, enter a name for the pipette. If the pipette has already been 

registered, then an informational message will appear.

Current connectivity status is always shown at the bottom left of the screen where it is 

clear if the pipette is:

 

USB connected 

 

Bluetooth connected 

If you want to connect to another pipette within the same channel as the first 

connected pipette, it will get disconnected and the new one will get connected while 

the sign below left remains the same.
For more information about all features PIPETMAN M Connected Software offers, 

please read the help section of PIPETMAN M Connected Software.

Chapter 8

PiPETTiNG SPEEd CONTrOL

PIPETMAN M Connected is set by default on speed 6 (maximum speed). You may need 

to change the speed of aspiration or dispensing depending on your application. You 

can change aspiration speed and dispensing speed independently: from very slow to 

very fast (speed 1 — speed 6). Your speed selection is memorized for each pipetting 

mode until you change it again.
1. 

Press the menu button to access the system menu, where you can find the 

different pipette modes, the speed menu, and configuration menu. 

2.  Select

 SPEEd 

by turning the push button, to confirm your selection click on  

the push button.

3.  Set 

ASPirATiON SPEEd: 

aspiration speed is blinking, set the speed by turning the 

push button and click to confirm your selection. 

4.  Set 

diSPENSiNG SPEEd: 

dispensing speed is blinking, set the speed by turning 

the push button and click to confirm your selection. 

After selecting the speed you go automatically back to pipetting mode.

Chapter 9

PiPETMAN® diAMONd TiPS

PIPETMAN DIAMOND Tips are made to the highest specifications; 

strict quality control is maintained throughout the manufacturing 

process. These tips are used to calibrate PIPETMAN M Connected, 

therefore for optimum performance we recommend using 

PIPETMAN DIAMOND Tips with your PIPETMAN M Connected. 
However, PIPETMAN M Connected also offers you a high 

compatibility with a large number of other tips. For more 

information, please contact your Gilson distributor.
Every PIPETMAN DIAMOND Tip is individually marked with an 

identification number (ID). To ensure accuracy and precision, 

Gilson’s quality assurance system focuses on the following critical 

parameters:

 

PIPETMAN DIAMOND Tips are made from pure polypropylene (virgin, 

metal-free, to avoid the possibility of contamination). They are available 

sterilized and with filters. 

 

Sterilized PIPETMAN DIAMOND Tips are certified free of detectable RNases, 

DNases, DNA, RNA, and proteases.

 

Optimized shape (revised collar for optimum sealing, thin walls, and fine point), 

making them easier to mount, more flexible, with no vortexing, and improved 

precision.

 

PIPETMAN DIAMOND Tips are free from even microscopic defects, especially 

at the orifice. All surfaces are smooth and hydrophobic, thereby avoiding the 

excessive retention of liquids that causes poor accuracy and a lack of precision.

 

ID Number

Figure 6

ID number

 

There is no connected 

pipette

Summary of Contents for F81040

Page 1: ...User s Guide EN...

Page 2: ...D 18 Tip Ejector Extension for Use with P10M 18 Personal Label Name and or Application Tag 19 GOOD LABORATORY PRACTICE FEATURES 19 General 19 Cycle Counters 19 Alarms 20 POWER MANAGEMENT 20 Battery Ch...

Page 3: ...fatigue and increase pipetting efficiency PIPETMAN M Connected is a unique pipette with only two buttons to operate the pipette and reach all menu options PIPETMAN M Connected features the classic PIP...

Page 4: ...9 10 11 12 13 14 15 16 17 18 19 Upper part Push button and volume adjustment knob Menu button direct access to all menu functions Tip ejector button Display Elastomer plug to cover the battery chargi...

Page 5: ...the tip by pressing the ejection button Your PIPETMAN M Connected is ready for the next pipetting cycle while maintaining the volume and the last settings Continue pipetting as you would with your mec...

Page 6: ...g protocol first change the pipetting speed of your pipette refer to Chapter 8 PIPETTING SPEED CONTROL on page 15 Aspiration can be done by using the standard speed by default speed 6 Dispensing shoul...

Page 7: ...nables personalized pipetting protocol creation on a PC or tablet for transfer to PIPETMAN M Connected through PIPETMAN M Connected Software an easy to use software To use this mode on a PC the softwa...

Page 8: ...pending on your application You can change aspiration speed and dispensing speed independently from very slow to very fast speed 1 speed 6 Your speed selection is memorized for each pipetting mode unt...

Page 9: ...4 5 P10mLM 5 7 4 5 P8x10M P12x10M 1 1 P8x20M P12x20M 2 3 1 P8x100M P12x100M 2 4 1 P8x200M P12x200M 2 3 1 P8x300M P12x300M 2 3 1 P8x1200M P12x1200M 2 4 2 3 Table 1 Immersion depth and wait time Mold an...

Page 10: ...surface of the liquid refer to Immersion Depth and Wait Time Table 1 on page 17 Change the tip before aspirating a different liquid sample or reagent Change the tip if a droplet remains at the end of...

Page 11: ...thium Ion battery Operation on the battery is done at user s risk only Opening the pipette voids the warranty Alarms Low battery warning Service is overdue warning refer to Chapter 15 MAINTENANCE on p...

Page 12: ...mber of readjustments carried out by this pipette NOTE Changes of service settings will reset all the service info to zero Chapter 14 CONFIGURATION Enter the menu using the menu button You will find n...

Page 13: ...tion Enter the sub menu in the same way and select the ADJUSTMENT menu The adjustment menu allows you to access the following items Standard readjustment Reset settings Standard Readjustment This menu...

Page 14: ...Chapter 21 REPLACEMENT PARTS on page 34 To access the O ring remove the tip ejector and unscrew the lower part of the tip holder You should now be able to remove the O ring from the piston Sometimes t...

Page 15: ...er 16 CLEANING AND DECONTAMINATION PIPETMAN M Connected is designed so that the parts normally in contact with contaminants can easily be cleaned and decontaminated As part of your quality system you...

Page 16: ...nents with distilled or sterile water 4 Leave the parts to dry by evaporation or wipe them with a clean lint free tissue or soft cloth 5 Lubricate the piston and reassemble the pipette according to th...

Page 17: ...e used with the recommended PIPETMAN DIAMOND Tips Under these conditions Gilson volumetric specifications in standard pipetting PIPET mode are guaranteed with a performance exceeding ISO 8655 2 recomm...

Page 18: ...22 F807022 F807022 F807022 F807022 F807022 F807022 Battery window F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005 Ejector window F807006 F807006 F807006 F807006 F807006 F807006 F807006...

Page 19: ...ture and pressure for distilled water TEMPERATURE AIR PRESSURE HPA C 800 853 907 960 1013 1067 15 1 0018 1 0018 1 0019 1 0019 1 0020 1 0020 15 5 1 0018 1 0019 1 0019 1 0020 1 0020 1 0021 16 1 0019 1 0...

Page 20: ...R IMPLIED INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL PUNITIVE INDIRECT OR IN...

Page 21: ...gilson com contactus LT801594 G 2021 Gilson Inc All rights reserved...

Reviews: