PIPETMAN® M CONNECTED
|
USER’S GUIDE
BENEFIT FROM THE BLUETOOTH® CONNECTION
|
PIPETMAN® M CONNECTED
Connection with the PIPETMAN M Connected Software
When it is necessary to have the pipette connected, the software will prompt you.
If the pipette has been registered previously, it is only necessary to activate the
Bluetooth on the pipette to make the connection; however, if this is the first time the
pipette is being used in the software, it is necessary to register the pipette.
1.
From the main menu, select
Other administration
and then
Pipette management
.
2. Select
next to
Registered pipettes
.
3. Connect the pipette.
If connecting via Bluetooth:
●
On the tablet, select
. The software searches for pipettes with Bluetooth
activated (indicator lights flashing).
1.
On the PC, select
Connect via BLE
. The software searches for pipettes with
Bluetooth activated (indicator lights flashing).
NOTE
Many devices may be detected, but only one can be connected.
If connecting via USB:
●
Use a USB cable to connect the pipette to the PC.
1.
Select
Connect via USB
. The connected pipette appears in the
Devices
list
2. Select the pipette in the
Devices
list.
3. Select
Next
.
4. When prompted, enter a name for the pipette. If the pipette has already been
registered, then an informational message will appear.
Current connectivity status is always shown at the bottom left of the screen where it is
clear if the pipette is:
●
USB connected
●
Bluetooth connected
If you want to connect to another pipette within the same channel as the first
connected pipette, it will get disconnected and the new one will get connected while
the sign below left remains the same.
For more information about all features PIPETMAN M Connected Software offers,
please read the help section of PIPETMAN M Connected Software.
Chapter 8
PiPETTiNG SPEEd CONTrOL
PIPETMAN M Connected is set by default on speed 6 (maximum speed). You may need
to change the speed of aspiration or dispensing depending on your application. You
can change aspiration speed and dispensing speed independently: from very slow to
very fast (speed 1 — speed 6). Your speed selection is memorized for each pipetting
mode until you change it again.
1.
Press the menu button to access the system menu, where you can find the
different pipette modes, the speed menu, and configuration menu.
2. Select
SPEEd
by turning the push button, to confirm your selection click on
the push button.
3. Set
ASPirATiON SPEEd:
aspiration speed is blinking, set the speed by turning the
push button and click to confirm your selection.
4. Set
diSPENSiNG SPEEd:
dispensing speed is blinking, set the speed by turning
the push button and click to confirm your selection.
After selecting the speed you go automatically back to pipetting mode.
Chapter 9
PiPETMAN® diAMONd TiPS
PIPETMAN DIAMOND Tips are made to the highest specifications;
strict quality control is maintained throughout the manufacturing
process. These tips are used to calibrate PIPETMAN M Connected,
therefore for optimum performance we recommend using
PIPETMAN DIAMOND Tips with your PIPETMAN M Connected.
However, PIPETMAN M Connected also offers you a high
compatibility with a large number of other tips. For more
information, please contact your Gilson distributor.
Every PIPETMAN DIAMOND Tip is individually marked with an
identification number (ID). To ensure accuracy and precision,
Gilson’s quality assurance system focuses on the following critical
parameters:
●
PIPETMAN DIAMOND Tips are made from pure polypropylene (virgin,
metal-free, to avoid the possibility of contamination). They are available
sterilized and with filters.
●
Sterilized PIPETMAN DIAMOND Tips are certified free of detectable RNases,
DNases, DNA, RNA, and proteases.
●
Optimized shape (revised collar for optimum sealing, thin walls, and fine point),
making them easier to mount, more flexible, with no vortexing, and improved
precision.
●
PIPETMAN DIAMOND Tips are free from even microscopic defects, especially
at the orifice. All surfaces are smooth and hydrophobic, thereby avoiding the
excessive retention of liquids that causes poor accuracy and a lack of precision.
ID Number
Figure 6
ID number
●
There is no connected
pipette